Far Cortical Locking Can Improve Healing of Fractures Stabilized with Locking Plates

被引:195
作者
Bottlang, Michael [1 ]
Lesser, Maren [2 ]
Koerber, Julia [3 ]
Doornink, Josef [1 ]
von Rechenberg, Brigitte [2 ]
Augat, Peter [3 ]
Fitzpatrick, Daniel C. [1 ]
Madey, Steven M. [1 ]
Marsh, J. Lawrence
机构
[1] Legacy Biomech Lab, Portland, OR 97232 USA
[2] Univ Zurich, Equine Dept, Musculoskeletal Res Unit, Vetsuisse Fac, CH-8057 Zurich, Switzerland
[3] Inst Biomech, D-82418 Murnau, Germany
基金
美国国家卫生研究院;
关键词
BIOMECHANICAL EVALUATION; TIBIA FRACTURES; LOCKED PLATES; FIXATION; BONE; SHEEP; OSTEOTOMY; STIFFNESS; MODEL; COMPRESSION;
D O I
10.2106/JBJS.I.01111
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
100224 [整形外科学];
摘要
Background: Locked bridge plating relies on secondary bone healing, which requires interfragmentary motion for callus formation. This study evaluated healing of fractures stabilized with a locked plating construct and a far cortical locking construct, which is a modified locked plating approach that promotes interfragmentary motion. The study tested whether far cortical locking constructs can improve fracture-healing compared with standard locked plating constructs. Methods: In an established ovine tibial osteotomy model with a 3-mm gap size, twelve osteotomies were randomly stabilized with locked plating or far cortical locking constructs applied medially. The far cortical locking constructs were designed to provide 84% lower stiffness than the locked plating constructs and permitted nearly parallel gap motion. Fracture-healing was monitored on weekly radiographs. After the animals were killed at week 9, healed tibiae were analyzed by computed tomography, mechanical testing in torsion, and histological examination. Results: Callus on weekly radiographs was greater in the far cortical locking constructs than in the locked plating constructs. At week 9, the far cortical locking group had a 36% greater callus volume (p = 0.03) and a 44% higher bone mineral content (p = 0.013) than the locked plating group. Callus in the locked plating specimens was asymmetric, having 49% less bone mineral content in the medial callus than in the lateral callus (p = 0.003). In far cortical locking specimens, medial and lateral callus had similar bone mineral content (p = 0.91). The far cortical locking specimens healed to be 54% stronger in torsion (p = 0.023) and sustained 156% greater energy to failure in torsion (p < 0.001) than locked plating specimens. Histologically, three of six locked plating specimens had deficient bridging across the medial cortex, while all remaining cortices had bridged. Conclusions: Inconsistent and asymmetric callus formation with locked plating constructs is likely due to their high stiffness and asymmetric gap closure. By providing flexible fixation and nearly parallel interfragmentary motion, far cortical locking constructs form more callus and heal to be stronger in torsion than locked plating constructs.
引用
收藏
页码:1652 / 1660
页数:9
相关论文
共 35 条
[1]
Refining animal models in fracture research:: seeking consensus in optimising both animal welfare and scientific validity for appropriate biomedical use [J].
Auer, Jorg A. ;
Goodship, Allen ;
Arnoczky, Steven ;
Pearce, Simon ;
Price, Jill ;
Claes, Lutz ;
Von Rechenberg, Brigitte ;
Hofmann-Amtenbrinck, Margarethe ;
Schneider, Erich ;
Mueller-Terpitz, R. ;
Thiele, F. ;
Rippe, Klaus-Peter ;
Grainger, David W. .
BMC MUSCULOSKELETAL DISORDERS, 2007, 8 (1)
[2]
Quantitative assessment of experimental fracture repair by peripheral computed tomography [J].
Augat, P ;
Merk, J ;
Genant, HK ;
Claes, L .
CALCIFIED TISSUE INTERNATIONAL, 1997, 60 (02) :194-199
[3]
Interfragmentary movement in diaphyseal tibia fractures fixed with locked intramedullary nails [J].
Augat, Peter ;
Penzkofer, Rainer ;
Nolte, Alexander ;
Maier, Michael ;
Panzer, Stefanie ;
Von Oldenburg, Geert ;
Pueschl, Klaus ;
Simon, Ulrich ;
Buehren, Volker .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2008, 22 (01) :30-36
[4]
Shear does not necessarifly inhibit bone healing [J].
Bishop, N. E. ;
van Rhijn, M. ;
Tami, I. ;
Corveleijn, R. ;
Schneider, E. ;
Ito, K. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2006, (443) :307-314
[5]
Far Cortical Locking Can Reduce Stiffness of Locked Plating Constructs While Retaining Construct Strength [J].
Bottlang, Michael ;
Doornink, Josef ;
Fitzpatrick, Daniel C. ;
Madey, Steven M. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2009, 91A (08) :1985-1994
[6]
Comparison of the mechanical performance of three types of external fixators: Linear, circular and hybrid [J].
Caja, VL ;
Kim, W ;
Larsson, S ;
Chao, EYS .
CLINICAL BIOMECHANICS, 1995, 10 (08) :401-406
[7]
ROLE OF INTERFRAGMENTARY STRAIN IN FRACTURE-HEALING - OVINE MODEL OF A HEALING OSTEOTOMY [J].
CHEAL, EJ ;
MANSMANN, KA ;
DIGIOIA, AM ;
HAYES, WC ;
PERREN, SM .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1991, 9 (01) :131-142
[8]
Influence of size and stability of the osteotomy gap on the success of fracture healing [J].
Claes, L ;
Augat, P ;
Suger, G ;
Wilke, HJ .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1997, 15 (04) :577-584
[9]
EFFECT OF DYNAMIZATION ON GAP HEALING OF DIAPHYSEAL FRACTURES UNDER EXTERNAL FIXATION [J].
CLAES, LE ;
WILKE, HJ ;
AUGAT, P ;
RUBENACKER, S ;
MARGEVICIUS, KJ .
CLINICAL BIOMECHANICS, 1995, 10 (05) :227-234
[10]
Claes LE, 1998, CLIN ORTHOP RELAT R, pS132