Simple method for the quantitative examination of extra column band broadening in microchromatographic systems

被引:13
作者
Beisler, AT [1 ]
Schaefer, KE [1 ]
Weber, SG [1 ]
机构
[1] Univ Pittsburgh, Dept Chem, Pittsburgh, PA 15260 USA
关键词
band broadening; post-column reactors;
D O I
10.1016/S0021-9673(02)02018-6
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
In recent years capillary chromatography has gained popularity for trace analyses. Most often UV or electrochemical detection is employed because the small peak volumes make post-column derivatization challenging. We have developed a simple method based on flow injection for determining contributions to peak broadening from post-column reactors. The only requirement for application of our methodology is that diffusion be in the Taylor regime so that radial concentration gradients are relaxed enabling mixing purely by diffusion. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:247 / 251
页数:5
相关论文
共 14 条
[1]  
BETTERIDGE D, 1978, ANAL CHEM, V50, pA832, DOI 10.1021/ac50031a001
[2]   Trace-level amino acid analysis by capillary liquid chromatography and application to in vivo microdialysis sampling with 10-s temporal resolution [J].
Boyd, BW ;
Witowski, SR ;
Kennedy, RT .
ANALYTICAL CHEMISTRY, 2000, 72 (04) :865-871
[3]   Automated capillary liquid chromatography for high sensitivity amino acid monitoring [J].
Boyd, BW ;
Kennedy, RT .
JOURNAL OF MICROCOLUMN SEPARATIONS, 2001, 13 (01) :24-32
[4]   Electrochemical detection of catecholamines at sub-5 fg levels by redox cycling [J].
Cullison, JK ;
Waraska, J ;
Buttaro, DJ ;
Acworth, IN ;
Bowers, ML .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1999, 19 (1-2) :253-259
[5]   ANALYSIS OF PEPTIDE MIXTURES BY CAPILLARY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY - A PRACTICAL GUIDE TO SMALL-SCALE SEPARATIONS [J].
DAVIS, MT ;
LEE, TD .
PROTEIN SCIENCE, 1992, 1 (07) :935-944
[6]  
Giddings J.C., 1991, Unified separation science
[7]   Determination of retinoids by reversed-phase capillary liquid chromatography with amperometric electrochemical detection [J].
Hagen, JJ ;
Washco, KA ;
Monnig, CA .
JOURNAL OF CHROMATOGRAPHY B-BIOMEDICAL APPLICATIONS, 1996, 677 (02) :225-231
[8]   THEORETICAL BACKGROUNDS OF FLOW-ANALYSIS [J].
HORVAI, G ;
PUNGOR, E .
CRC CRITICAL REVIEWS IN ANALYTICAL CHEMISTRY, 1987, 17 (03) :231-264
[9]  
Kakuta M, 2001, CHEM REC, V1, P395
[10]  
PIPES LA, 1970, APPL MATH ENG PHYSIC