A study of temperature-modulated differential scanning calorimetry with high-resolution infrared thermography

被引:18
作者
Androsch, R
Pyda, M
Wang, H
Wunderlich, B [1 ]
机构
[1] Univ Tennessee, Dept Chem, Knoxville, TN 37996 USA
[2] Oak Ridge Natl Lab, High Temp Mat Lab, Oak Ridge, TN 37831 USA
[3] Oak Ridge Natl Lab, Div Chem & Analyt Sci, Oak Ridge, TN 37831 USA
基金
美国国家科学基金会;
关键词
differential scanning calorimetry (DSC); infrared thermography; phase angle; temperature gradient; temperature-modulated differential scanning calorimetry (TMDSC);
D O I
10.1023/A:1010158621847
中图分类号
O414.1 [热力学];
学科分类号
摘要
Temperature gradients within the sample and furnace in temperature-modulated differential scanning calorimetry (TMDSC) are studied for a power-compensation calorimeter of the Perkin Elmer type. The temperature measurements were made with a high-speed, high-resolution infrared camera. Differences between programmed and actual temperature amplitudes are determined as a function of sample thickness for a sawtooth modulation with up to 48 K min(-1) heating and cooling rates. Phase angles have been established, and the effect of open and sealed sample pans has been analyzed. A simple one-dimensional description of the observed effects is made and a three-dimensional one is suggested based on a model available in the literature.
引用
收藏
页码:661 / 679
页数:19
相关论文
共 22 条
[21]  
WUNDERLICH B, 1997, THERMAL CHARACTERIZA, V1, P205
[22]  
WUNDERLICH B, 1971, PHYSICAL METHODS C 5, V1, P427