Vitro examinations for detection of vascular stenoses using spiral CT angiography.

被引:2
作者
Beilicke, M [1 ]
Kloppel, R [1 ]
Lieberenz, S [1 ]
机构
[1] Univ Leipzig, Radiol Diagnost Klin, D-04103 Leipzig, Germany
来源
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN | 1998年 / 168卷 / 03期
关键词
spiral CT angiography; stenoses;
D O I
10.1055/s-2007-1015116
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: This experimental study investigates the suitability of spiral CT angiography for detection of stenoses. Material and methods: Three vessel phantoms of 4, 7, and 10 mm diameter were scanned with spiral CT several times using different scanning parameters each time. Axial scans were viewed in different ways. Stenoses were adjusted axial, oblique, and parallel to the scan plane, and in between. 7 radiologists had to make a total of 630 decisions about the detectability of stenoses. Results: Adequate scan- and reconstructional parameters improve detection of stenoses significantly. 100% recognition can be reached with 5 mm layer thickness and table feed and 2 mm reconstructional interval provided axial scans are viewed in cine mode on a monitor. Vessels running parallel to the layer plane show better results than vessels axial to the plane, which in turn show better results than those running oblique. Vessels of 4 to 10 mm diameter are almost equally suitable. Conclusions: Viewing axial scans, the detectability is rather influenced by different locations of stenoses than by varying vessel diameters.
引用
收藏
页码:217 / 221
页数:5
相关论文
共 11 条
[1]   TECHNICAL OPTIMIZATION OF SPIRAL CT FOR DEPICTION OF RENAL-ARTERY STENOSIS - IN-VITRO ANALYSIS [J].
BRINK, JA ;
LIM, JT ;
WANG, G ;
HEIKEN, JP ;
DEYOE, LA ;
VANNIER, MW .
RADIOLOGY, 1995, 194 (01) :157-163
[2]   CAROTID-ARTERY STENOSIS - A PROSPECTIVE COMPARISON OF CT ANGIOGRAPHY AND CONVENTIONAL ANGIOGRAPHY [J].
CUMMING, MJ ;
MORROW, IM .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1994, 163 (03) :517-523
[3]   CT ANGIOGRAPHY - APPLICATION TO THE EVALUATION OF CAROTID-ARTERY STENOSIS [J].
DILLON, EH ;
VANLEEUWEN, MS ;
FERNANDEZ, MA ;
EIKELBOOM, BC ;
MALI, WPTM .
RADIOLOGY, 1993, 189 (01) :211-219
[4]   RENAL ARTERIAL STENOSES - SPIRAL CT ANGIOGRAPHY [J].
GALANSKI, M ;
PROKOP, M ;
CHAVAN, A ;
SCHAEFER, CM ;
JANDELEIT, K ;
NISCHELSKY, JE .
RADIOLOGY, 1993, 189 (01) :185-192
[5]   DIAGNOSIS OF CAROTID-ARTERY DISEASE - PRELIMINARY EXPERIENCE WITH MAXIMUM-INTENSITY-PROJECTION SPIRAL CT ANGIOGRAPHY [J].
MARKS, MP ;
NAPEL, S ;
JORDAN, JE ;
ENZMANN, DR .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1993, 160 (06) :1267-1271
[6]   BLINDED-READER COMPARISON OF MAGNETIC-RESONANCE ANGIOGRAPHY AND DUPLEX ULTRASONOGRAPHY FOR CAROTID-ARTERY BIFURCATION STENOSIS [J].
MITTL, RL ;
BRODERICK, M ;
CARPENTER, JP ;
GOLDBERG, HI ;
LISTRUD, J ;
MISHKIN, MM ;
BERKOWITZ, HD ;
ATLAS, SW .
STROKE, 1994, 25 (01) :4-10
[7]   CT ANGIOGRAPHY AND ARTERIAL DSA IN ASSESSING OCCLUSIVE PROCESSES OF THE PELVIC ARTERIES - 1ST RESULTS [J].
RICHTER, CS ;
BIAMINO, G ;
NIEMANN, VT ;
RAGG, C ;
FELIX, R .
FORTSCHRITTE AUF DEM GEBIETE DER RONTGENSTRAHLEN UND DER NEUEN BILDGEBENDEN VERFAHREN, 1994, 161 (02) :154-160
[8]   SPIRAL CT OF RENAL-ARTERY STENOSIS - COMPARISON OF 3-DIMENSIONAL RENDERING TECHNIQUES [J].
RUBIN, GD ;
DAKE, MD ;
NAPEL, S ;
JEFFREY, RB ;
MCDONNELL, CH ;
SOMMER, FG ;
WEXLER, L ;
WILLIAMS, DM .
RADIOLOGY, 1994, 190 (01) :181-189
[9]   COMMON CAROTID-ARTERY BIFURCATION - EVALUATION WITH SPIRAL CT - WORK IN PROGRESS [J].
SCHWARTZ, RB ;
JONES, KM ;
CHERNOFF, DM ;
MUKHERJI, SK ;
KHORASANI, R ;
TICE, HM ;
KIKINIS, R ;
HOOTON, SM ;
STIEG, PE ;
POLAK, JF .
RADIOLOGY, 1992, 185 (02) :513-519
[10]  
WHITE JE, 1994, AM SURGEON, V60, P340