Clinical evaluation of a new high efficiency hemodialyzer:: Polysynthane™

被引:14
作者
Clark, WR
Shinaberger, JH
机构
[1] Baxter Healthcare Corp, Div Renal, Mcgaw Pk, IL USA
[2] Indiana Univ, Sch Med, Div Nephrol, Indianapolis, IN USA
[3] W Los Angeles Dept Vet Affairs Med Ctr, Dialysis Unit, Los Angeles, CA USA
关键词
D O I
10.1097/00002480-200005000-00009
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Two increasingly common characteristics of the American chronic hemodialysis (HD) population, high hematocrit and large body size, may vender the currently recommended adequacy targets difficult to achieve, even with very efficient dialyzers. In a group of patients with these characteristics, we assessed the ability of a new high efficiency dialyzer (PSN(TM)210; Baxter Healthcare Corporation) to achieve the currently recommended adequacy targets. Six patients (mean pre-HD weight and hematocrit, 90.3 +/- 18.0 kg and 0.40 +/- 0.03 kg, respectively) were evaluated. At prescribed blood and dialysate flow rates of 400 and 800 ml/min, respectively, and a mean treatment duration of 4 hrs, mean delivered urea Kt/V and reduction ratio (URR) were 1.38 +/- 0.25 and 0.73 +/- 0.07, respectively. For the same flow rates, whole blood clearances for urea, creatinine, and phosphate were 315 +/- 13, 246 +/- 28, and 260 +/- 27 ml/min, respectively. These data indicate this dialyzer has an efficient mass transfer design allowing adequate dialysis to be delivered even to very large patients under high efficiency conditions.
引用
收藏
页码:288 / 292
页数:5
相关论文
共 30 条
[11]   A MECHANISTIC ANALYSIS OF THE NATIONAL COOPERATIVE DIALYSIS STUDY (NCDS) [J].
GOTCH, FA ;
SARGENT, JA .
KIDNEY INTERNATIONAL, 1985, 28 (03) :526-534
[12]  
Guenther C., 1999, Nephrology Dialysis Transplantation, V14, pA209
[13]   EFFECTS OF DOSE OF DIALYSIS ON MORBIDITY AND MORTALITY [J].
HAKIM, RM ;
BREYER, J ;
ISMAIL, N ;
SCHULMAN, G .
AMERICAN JOURNAL OF KIDNEY DISEASES, 1994, 23 (05) :661-669
[14]   CONVECTIVE MASS-TRANSFER IN HEMODIALYSIS [J].
JAFFRIN, MY .
ARTIFICIAL ORGANS, 1995, 19 (11) :1162-1171
[15]  
KESHAVIAH P, 1986, ARTIF ORGANS, V110, P189
[16]   Gender-specific differences in dialysis quality (Kt/V):: 'big men' are at risk of inadequate haemodialysis treatment [J].
Kuhlmann, MK ;
König, J ;
Riegel, W ;
Köhler, H .
NEPHROLOGY DIALYSIS TRANSPLANTATION, 1999, 14 (01) :147-153
[17]   Hemodialyzer mass transfer-area coefficients for urea increase at high dialysate flow rates [J].
Leypoldt, JK ;
Cheung, AK ;
Agodoa, LY ;
Daugirdas, JT ;
Greene, T ;
Keshaviah, PR .
KIDNEY INTERNATIONAL, 1997, 51 (06) :2013-2017
[18]  
LEYPOLDT JK, IN PRESS HOME HD INT
[19]  
LIPPS BJ, 1967, T AM SOC ART INT ORG, V13, P200
[20]   EFFECT OF THE HEMODIALYSIS PRESCRIPTION ON PATIENT MORBIDITY - REPORT FROM THE NATIONAL COOPERATIVE DIALYSIS STUDY [J].
LOWRIE, EG ;
LAIRD, NM ;
PARKER, TF ;
SARGENT, JA .
NEW ENGLAND JOURNAL OF MEDICINE, 1981, 305 (20) :1176-1181