Semicontinuous intra-abdominal pressure measurement using an intragastric Compliance catheter

被引:20
作者
De Waele, Jan J. [1 ]
Berrevoet, F. [1 ]
Reyntjens, K. [1 ]
Pletinckx, P. [1 ]
De laet, I. [1 ]
Hoste, E. [1 ]
机构
[1] State Univ Ghent Hosp, Intens Care Unit, B-9000 Ghent, Belgium
关键词
intra-abdominal pressure; intra-abdominal hypertension; abdominal compartment syndrome; surgery; trauma; critically ill patients;
D O I
10.1007/s00134-007-0682-3
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: To compare intra-abdominal pressure (IAP) measurements obtained from an intragastric Compliance catheter with the pressure measured directly in the abdominal cavity. Design and setting: Prospective cohort study in an operating room of the Ghent University Hospital Patients: Seven patients undergoing elective laparoscopic cholecystectomy. Interventions: IAP was obtained from both an intragastric catheter and directly from the peritoneal cavity at 1-minute intervals in patients undergoing elective cholecystectomy and compared using Bland-Altman analysis. Measurements and results: In 156 paired measurements obtained from 7 patients the mean difference between IAP(gastric) and IAP(ref) was 0.12 +/- 0.70mmHg (95% CI 0.01-0.23). Conclusions: IAP measured using an intragastric Compliance catheter reliably reflects the reference IAP in patients undergoing laparoscopic cholecystectomy.
引用
收藏
页码:1297 / 1300
页数:4
相关论文
共 13 条
[1]   Continuous intra-abdominal pressure measurement technique [J].
Balogh, Z ;
Jones, F ;
D'Amours, S ;
Parr, M ;
Sugrue, M .
AMERICAN JOURNAL OF SURGERY, 2004, 188 (06) :679-683
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   Validation of direct intraabdominal pressure measurement using a continuous indwelling compartment pressure monitor [J].
Brooks, AJ ;
Simpson, A ;
Delbridge, M ;
Beckingham, IJ ;
Girling, KJ .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2005, 58 (04) :830-832
[4]   BEDSIDE MEASUREMENT OF INTRAABDOMINAL PRESSURE (IAP) VIA AN INDWELLING NASO-GASTRIC TUBE - CLINICAL VALIDATION OF THE TECHNIQUE [J].
COLLEE, GG ;
LOMAX, DM ;
FERGUSON, C ;
HANSON, GC .
INTENSIVE CARE MEDICINE, 1993, 19 (08) :478-480
[5]   Intra- and interobserver variability during in vitro validation of two novel methods for intra-abdominal pressure monitoring [J].
De Potter, TJR ;
Dits, H ;
Malbrain, MLNG .
INTENSIVE CARE MEDICINE, 2005, 31 (05) :747-751
[6]   Fluid vs. air for semicontinuous intra-abdominal pressure measurements using a compliance catheter [J].
De Waele, J ;
Billiet, EAI ;
Hoste, E ;
Blot, SI ;
Colardyn, FA .
INTENSIVE CARE MEDICINE, 2005, 31 (04) :598-599
[7]   Gastric tonometry and direct intraabdominal pressure monitoring in abdominal compartment syndrome [J].
Engum, SA ;
Kogon, B ;
Jensen, E ;
Isch, J ;
Balanoff, C ;
Grosfeld, JL .
JOURNAL OF PEDIATRIC SURGERY, 2002, 37 (02) :214-217
[8]   Estimation of intra-abdominal pressure by bladder pressure measurement: Validity and methodology [J].
Fusco, MA ;
Martin, RS ;
Chang, MC .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2001, 50 (02) :297-302
[9]   Abdominal compartment syndrome: does intra-cystic pressure reflect actual intra-abdominal pressure? A prospective study in surgical patients [J].
Johna, S ;
Taylor, E ;
Brown, C ;
Zimmerman, G .
CRITICAL CARE, 1999, 3 (06) :135-138
[10]   Prevalence of intra-abdominal hypertension in critically ill patients: a multicentre epidemiological study [J].
Malbrain, MLNG ;
Chiumello, D ;
Pelosi, P ;
Wilmer, A ;
Brienza, N ;
Malcangi, V ;
Bihari, D ;
Innes, R ;
Cohen, J ;
Singer, P ;
Japiassu, A ;
Kurtop, E ;
De Keulenaer, BL ;
Daelemans, R ;
Del Turco, M ;
Cosimini, P ;
Ranieri, M ;
Jacquet, L ;
Laterre, PF ;
Gattinoni, L .
INTENSIVE CARE MEDICINE, 2004, 30 (05) :822-829