Towards a button-sized 1024-site wireless cortical microstianulating array

被引:22
作者
Ghovanloo, M [1 ]
Wise, KD [1 ]
Najafi, K [1 ]
机构
[1] Univ Michigan, Ctr Wireless Integrated Microsyst, Ann Arbor, MI 48109 USA
来源
1ST INTERNATIONAL IEEE EMBS CONFERENCE ON NEURAL ENGINEERING 2003, CONFERENCE PROCEEDINGS | 2003年
关键词
microelectrode; wireless; implant; stimulator; cortex;
D O I
10.1109/CNE.2003.1196776
中图分类号
TP18 [人工智能理论];
学科分类号
081104 ; 0812 ; 0835 ; 1405 ;
摘要
Design and development of a micromachined 16-site wireless addressable active stimulating microprobe, called Interestim-3, based on the University of Michigan MEMS 3-D microelectrode-array technology have been discussed. Up to 64 Interestim-3 probes can be used in parallel with only 2 connections to a mutual hybrid coil in order to achieve a 1024-site intracortical microstimulating system with no trans or subcutaneous wires. 3-D microassembly of Interestim-3 probes, which is facilitated by small number of interconnects (2/probe) that also gives way to a high yield has been shown. Interestim-3 has been fabricated in the UM 3-mum 1M/2P BiCMOS process. It has four 60mum x 3.1mm shanks with 250mum spacing, each having four 181mum(2) sites. Interestim-3 circuitry has about 4200 transistors that occupy 21.2 mm(2) of 32.5 mm(2) probe back-end.
引用
收藏
页码:138 / 141
页数:4
相关论文
共 13 条
[1]   A high-yield microassembly structure for three-dimensional microelectrode arrays [J].
Bai, Q ;
Wise, KD ;
Anderson, DJ .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2000, 47 (03) :281-289
[2]   VISUAL SENSATIONS PRODUCED BY INTRACORTICAL MICROSTIMULATION OF THE HUMAN OCCIPITAL CORTEX [J].
BAK, M ;
GIRVIN, JP ;
HAMBRECHT, FT ;
KUFTA, CV ;
LOEB, GE ;
SCHMIDT, EM .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1990, 28 (03) :257-259
[3]   A SILICON-BASED, 3-DIMENSIONAL NEURAL INTERFACE - MANUFACTURING PROCESSES FOR AN INTRACORTICAL ELECTRODE ARRAY [J].
CAMPBELL, PK ;
JONES, KE ;
HUBER, RJ ;
HORCH, KW ;
NORMANN, RA .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1991, 38 (08) :758-768
[4]   PHOSPHENES PRODUCED BY ELECTRICAL-STIMULATION OF HUMAN OCCIPITAL CORTEX, AND THEIR APPLICATION TO DEVELOPMENT OF A PROSTHESIS FOR BLIND [J].
DOBELLE, WH ;
MLADEJOVSKY, MG .
JOURNAL OF PHYSIOLOGY-LONDON, 1974, 243 (02) :553-+
[5]   A BiCMOS wireless stimulator chip for micromachined stimulating microprobes [J].
Ghovanloo, M ;
Najafi, K .
SECOND JOINT EMBS-BMES CONFERENCE 2002, VOLS 1-3, CONFERENCE PROCEEDINGS: BIOENGINEERING - INTEGRATIVE METHODOLOGIES, NEW TECHNOLOGIES, 2002, :2113-2114
[6]   Fully integrated power supply design for wireless biomedical implants [J].
Ghovanloo, M ;
Najafi, K .
2ND ANNUAL INTERNATIONAL IEEE-EMBS SPECIAL TOPIC CONFERENCE ON MICROTECHNOLOGIES IN MEDICINE & BIOLOGY, PROCEEDINGS, 2002, :414-419
[7]   A BiCMOS wireless interface chip for micromachined stimulating microprobes [J].
Ghovanloo, M ;
Beach, K ;
Wise, KD ;
Najafi, K .
2ND ANNUAL INTERNATIONAL IEEE-EMBS SPECIAL TOPIC CONFERENCE ON MICROTECHNOLOGIES IN MEDICINE & BIOLOGY, PROCEEDINGS, 2002, :277-282
[8]  
GHOVANLOO M, 2002, IEEE 45 MIDW S CIRC
[9]  
GINGERICH MD, 2002, THESIS U MICHIGAN AN
[10]   A 64-site multishank CMOS low-profile neural stimulating probe [J].
Kim, C ;
Wise, KD .
IEEE JOURNAL OF SOLID-STATE CIRCUITS, 1996, 31 (09) :1230-1238