19 mm sized bileaflet valve prostheses' flow field investigated by bidimensional laser Doppler anemometry (part II: maximum turbulent shear stresses)

被引:18
作者
Barbaro, V [1 ]
Grigioni, M [1 ]
Daniele, C [1 ]
D'Avenio, G [1 ]
Boccanera, G [1 ]
机构
[1] Ist Super Sanita, Lab Ingn Biomed, I-00161 Rome, Italy
关键词
flow visualization; laser Doppler anemometry; aortic cardiac valve prosthesis; in vitro testing; peak turbulent shear stress;
D O I
10.1177/039139889702001106
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The investigation of the flow field generated by cardiac valve prostheses is a necessary task to gain knowledge on the possible relationship between turbulence-derived stresses and the hemolytic and thrombogenic complications in patients after valve replacement. The study of turbulence flows downstream of cardiac prostheses, in literature, especially concerns large-sized prostheses with a variable flow regime from very low up to 6 L/min. The Food and Drug Administration draft guidance requires the study of the minimum prosthetic size at a high cardiac output to reach the maximum Reynolds number conditions. Within the framework of a national research project regarding the characterization of cardiovascular endoprostheses, an in-depth study of turbulence generated downstream of bileaflet cardiac valves is currently under way at the Laboratory of Biomedical Engineering of the Istituto Superiore di Sanita. Four models of 19 mm bileaflet valve prostheses were used: St Jude Medical HF: Edwards Tekna, Sorin Bicarbon, and CarbonMedics. The prostheses were selected for the nominal Tissue Annulus Diameter as reported by manufacturers without any assessment of valve sizing method, and were mounted in aortic position. The aortic geometry was scaled for 19 mm prostheses using angiographic data. The turbulence-derived shear stresses were investigated very close to the valve (0.35 D-0), using a bidimensional Laser Doppler anemometry system and applying the Principal Stress Analysis. Results concern typical turbulence quantities during a 50 ms window at peak flow in the systolic phase. Conclusions are drawn regarding the turbulence associated to valve design features, as well as the possible damage to blood constituents.
引用
收藏
页码:629 / 636
页数:8
相关论文
共 20 条
[1]   DETERMINATION OF PRINCIPAL REYNOLDS STRESSES IN PULSATILE FLOWS AFTER ELLIPTIC FILTERING OF DISCRETE VELOCITY-MEASUREMENTS [J].
BALDWIN, JT ;
DEUTSCH, S ;
PETRIE, HL ;
TARBELL, JM .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1993, 115 (04) :396-403
[2]   19 mm sized bileaflet valve prostheses' flow field investigated by bidimensional laser Doppler anemometry (part I: velocity profiles) [J].
Barbaro, V ;
Grigioni, M ;
Daniele, C ;
D'Avenio, G ;
Boccanera, G .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1997, 20 (11) :622-628
[3]   APPROACHING COMPARABILITY AND RESULTS OF PULSATILE FLOW IN-VITRO TESTING OF PROSTHETIC HEART-VALVES [J].
BARBARO, V ;
BOCCANERA, G ;
DANIELE, C ;
GRIGIONI, M ;
PALOMBO, A .
JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1995, 19 (04) :115-118
[4]  
BARBARO V, 1997, 4 ESEM C 25 28 MAY 1
[5]  
Blackshear P., 1972, Chemistry of Biosurfaces, P523
[6]  
*FDA, 1994, ITS DRAFT REPL HEART
[7]  
Fisher J, 1994, J Heart Valve Dis, V3, P75
[8]   Identification of peak stresses in cardiac prostheses - A comparison of two-dimensional versus three-dimensional principal stress analyses [J].
Fontaine, AA ;
Ellis, JT ;
Healy, TM ;
Hopmeyer, J ;
Yoganathan, AP .
ASAIO JOURNAL, 1996, 42 (03) :154-163
[9]   ESTIMATION OF SHEAR STRESS-RELATED BLOOD DAMAGE IN HEART-VALVE PROSTHESES - INVITRO COMPARISON OF 25 AORTIC VALVES [J].
GIERSIEPEN, M ;
WURZINGER, LJ ;
OPITZ, R ;
REUL, H .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1990, 13 (05) :300-306
[10]   VELOCITY AND SHEAR-STRESS DISTRIBUTION DOWNSTREAM OF MECHANICAL HEART-VALVES IN PULSATILE FLOW [J].
GIERSIEPEN, M ;
KRAUSE, U ;
KNOTT, E ;
REUL, H ;
RAU, G .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1989, 12 (04) :261-269