Low contrast imaging with a GaAs pixel digital detector

被引:20
作者
Amendolia, SR
Bisogni, MG
Bottigli, U
Ciocci, MA
Delogu, P
Dipasquale, G
Fantacci, ME
Giannelli, M
Maestro, P
Marzulli, VM
Pernigotti, E
Rosso, V
Stefanini, A
Stumbo, S
机构
[1] Univ Sassari, Dipartimento Matemat & Fis, I-56010 Pisa, Italy
[2] Sez INFN, Pisa, Italy
[3] Univ Pisa, Dipartimento Fis, Pisa, Italy
[4] Sez INFN, Pisa, Italy
[5] Univ Siena, Dipartimento Fis, I-53100 Siena, Italy
[6] Sez INFN, Pisa, Italy
[7] Univ Pisa, Dipartimento Fis, Pisa, Italy
[8] INFM, Unita Ric, Pisa, Italy
关键词
D O I
10.1109/23.873000
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A digital mammography system based on GaAs pixel detector has been developed by the INFN (Istituto Nazionale di Fisica Nucleare) collaboration MED46. The high atomic number makes the GaAs a very efficient material for low energy X-rays detection (10 - 30 keV is the typical energy range used in mammography). Low contrast details can be detected with a significant dose reduction to the patient. The system presented in this paper consists of a 4096 pixel matrix built on a 200 mu m thick Semi Insulating GaAs substrate. The pixel size is 170 x 170 mu m(2) for a total active area of 1.18 cm(2). The detector is bump-bonded to a VLSI Front-End:chip which implements a single-photon counting architecture. This feature allows to enhance the radiographic contrasts detection with respect to charge integrating devices. The system has been tested by using a standard mammographic tube. Images of mammographic phantoms will be presented and compared with radiographs obtained with traditional :film/screen systems. Monte Carlo simulations have been also performed to evaluate the imaging capability of the system. Comparison with simulations and experimental results will be shown.
引用
收藏
页码:1478 / 1482
页数:5
相关论文
共 19 条
[1]  
*3M INC, 3M MED IM FILM M TRI
[2]   PERFORMANCE OF A NEW OHMIC CONTACT FOR GAAS PARTICLE DETECTORS [J].
ALIETTI, M ;
CANALI, C ;
CASTALDINI, A ;
CAVALLINI, A ;
CETRONIO, A ;
CHIOSSI, C ;
DAURIA, S ;
DELPAPA, C ;
LANZIERI, C ;
NAVA, F ;
VANNI, P .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1995, 362 (2-3) :344-348
[3]  
*AM CANC SOC, 1994, CANC FACTS FIG 1994
[4]   MEDIPIX: a VLSI chip for a GaAs pixel detector for digital radiology [J].
Amendolia, SR ;
Bertolucci, E ;
Bisogni, MG ;
Bottigli, U ;
Ceccopieri, A ;
Ciocci, MA ;
Conti, M ;
Delogu, P ;
Fantacci, ME ;
Maestro, P ;
Marzulli, V ;
Pernigotti, E ;
Romeo, N ;
Rosso, V ;
Rosso, P ;
Stefanini, A ;
Stumbo, S .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1999, 422 (1-3) :201-205
[5]  
Amendolia SR, 1999, NUOVO CIMENTO A, V112, P167
[6]   Experimental study of LEC GaAs detectors for X-ray digital radiography [J].
Amendolia, SR ;
Bisogni, MG ;
Campbell, M ;
Cola, A ;
DAuria, S ;
DaVia, C ;
Heijne, EHM ;
Fantacci, ME ;
OShea, V ;
Rosso, V ;
Smith, K ;
Vasanelli, L .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1996, 380 (1-2) :410-413
[7]  
Amendolia SR, 1998, PHYS MEDICA, V14, P17
[8]  
Amendolia SR, 1997, PHYS MEDICA, V13, P157
[9]   EVALUATION OF ELEMENTAL AND COMPOUND SEMICONDUCTORS FOR X-RAY DIGITAL RADIOGRAPHY [J].
BENCIVELLI, W ;
BERTOLUCCI, E ;
BOTTIGLI, U ;
DELGUERRA, A ;
MESSINEO, A ;
NELSON, WR ;
RANDACCIO, P ;
ROSSO, V ;
RUSSO, P ;
STEFANINI, A .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1991, 310 (1-2) :210-214
[10]  
BIELAJEW AF, 1987, NUCL INSTRUM METH B, V18, P165