In-column field-amplified sample stacking of biogenic amines on microfabricated ellectrophoresis devices

被引:50
作者
Beard, NR
Zhang, CX
deMello, AJ [1 ]
机构
[1] Univ London Imperial Coll Sci Technol & Med, Dept Chem, London SW7 2AY, England
[2] NPS Pharmaceut, Salt Lake City, UT USA
关键词
biogenic amines; electrophoresis; field-amplified sample stacking; microchip; sample preconcentration;
D O I
10.1002/elps.200390088
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A novel method for performing in-column field-amplified sample stacking (FASS) in chip-based electrophoretic systems is presented. The methodology involves the use of a narrow sample channel (NSC) injector. NSC injectors allow sample plugs to be introduced directly into the separation channel, and subsequent stacking and separation can proceed without any need for leakage control. More importantly, stacking and separation occur in a single step negating the requirement for complex channel geometries and voltage switching to control sample plugs during the stacking procedure. The chip is composed of six paralleled systems. Using the NSC injector design, the number of reservoirs in the multiplexed chip is reduced to N + 2, where N is the number of paralleled systems. This design feature radically reduces the complexity in chip structures and associated chip operation. The approach is applied to the analysis of fluorescently labelled biogenic amines affording detection at concentrations down to 20 pM.
引用
收藏
页码:732 / 739
页数:8
相关论文
共 44 条
[1]  
Beard NP, 2002, ELECTROPHORESIS, V23, P1722, DOI 10.1002/1522-2683(200206)23:11<1722::AID-ELPS1722>3.0.CO
[2]  
2-W
[3]   OPTIMIZATION IN SAMPLE STACKING FOR HIGH-PERFORMANCE CAPILLARY ELECTROPHORESIS [J].
BURGI, DS ;
CHIEN, RL .
ANALYTICAL CHEMISTRY, 1991, 63 (18) :2042-2047
[4]   Application of capillary electrophoresis with laser-induced fluorescence detection to the determination of biogenic amines and amino acids in brain microdialysate and homogenate samples [J].
Chen, Z ;
Wu, J ;
Baker, GB ;
Parent, M ;
Dovichi, NJ .
JOURNAL OF CHROMATOGRAPHY A, 2001, 914 (1-2) :293-298
[5]   ELECTROOSMOTIC PROPERTIES AND PEAK BROADENING IN FIELD-AMPLIFIED CAPILLARY ELECTROPHORESIS [J].
CHIEN, RL ;
HELMER, JC .
ANALYTICAL CHEMISTRY, 1991, 63 (14) :1354-1361
[6]   SAMPLE STACKING OF AN EXTREMELY LARGE INJECTION VOLUME IN HIGH-PERFORMANCE CAPILLARY ELECTROPHORESIS [J].
CHIEN, RL ;
BURGI, DS .
ANALYTICAL CHEMISTRY, 1992, 64 (09) :1046-1050
[7]   ON-COLUMN TRANSIENT AND COUPLED COLUMN ISOTACHOPHORETIC PRECONCENTRATION OF PROTEIN SAMPLES IN CAPILLARY ZONE ELECTROPHORESIS [J].
FORET, F ;
SZOKO, E ;
KARGER, BL .
JOURNAL OF CHROMATOGRAPHY, 1992, 608 (1-2) :3-12
[8]   INDIRECT FLUORIMETRIC DETECTION AND QUANTIFICATION IN CAPILLARY ZONE ELECTROPHORESIS OF INORGANIC ANIONS AND NUCLEOTIDES [J].
GROSS, L ;
YEUNG, ES .
JOURNAL OF CHROMATOGRAPHY, 1989, 480 :169-178
[9]   Large-volume sample stacking in acidic buffer for analysis of small organic and inorganic anions by capillary electrophoresis [J].
He, Y ;
Lee, HK .
ANALYTICAL CHEMISTRY, 1999, 71 (05) :995-1001
[10]   Microchip structures for submillisecond electrophoresis [J].
Jacobson, SC ;
Culbertson, CT ;
Daler, JE ;
Ramsey, JM .
ANALYTICAL CHEMISTRY, 1998, 70 (16) :3476-3480