Point of care monitoring of oral anticoagulant therapy in children:: comparison of CoaguChek Plus and Thrombotest methods with venous international normalised ratio

被引:26
作者
Ignjatovic, V
Barnes, C
Newall, F
Hamilton, S
Burgess, J
Monagle, P
机构
[1] Royal Childrens Hosp, Dept Haematol, Parkville, Vic 3143, Australia
[2] Royal Childrens Hosp, Core Lab, Parkville, Vic 3143, Australia
[3] Royal Childrens Hosp, Pathol Collect Dept, Parkville, Vic 3143, Australia
关键词
anticoagulant monitoring; children; International Normalised Ratio (INR); Point of Care (POC); warfarin;
D O I
10.1160/TH04-02-0085
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
This paper reports the outcome of a research protocol aimed at optimising warfarin monitoring in a tertiary pediatric centre. The Thrombotest(TM) INR was the standard monitoring test employed to manage oral anticoagulant therapy in children at the Royal Children's Hospital (RCH), Melbourne. This study compares the results of this standard method to the novel CoaguChek(R) INR monitor and the "gold standard" technique of venous INR sampling. The objectives were to determine 1) if point-of-care techniques of measuring the INR (Thrombotest and CoaguChek) are accurate and reliable compared to INR results obtained from venous sampling, processed in an accredited laboratory, and 2) if INR results generated by POC devices can be safely used to manage oral anticoagulant therapy in children. 18 children (10 females and 8 males) participated in the study. Ages ranged from 9 months to 21 years (Mean 11.9 years; SD 5.03 years). The agreement between CoaguChek and venous INR measurements (r = 0.885) was shown to be higher compared to Thrombotest and venous INR (r = 0.700). Compared to the venous INR, values obtained with Coaguchek and Thrombotest crossed into or out of the therapeutic range in 25% and 36% of cases respectively. In 88% of the CoaguChek cases and 57% Thrombotest cases, the difference from the venous result was less than 0.5. The CoaguChek method of INR monitoring is a more accurate and reliable method compared to Thrombotest, in the pediatric population tested, and can be safely used to manage oral anticoagulant therapy in children.
引用
收藏
页码:734 / 737
页数:4
相关论文
共 23 条
[1]  
ANDREW M, 2000, COMPLICATIONS INFANC
[2]  
BELSEY RE, 1991, J FAM PRACTICE, V33, P266
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   Accuracy, clinical correlation, and patient acceptance of two handheld prothrombin time monitoring devices in the ambulatory setting [J].
Chapman, DC ;
Stephens, MA ;
Hamann, GL ;
Bailey, LE ;
Dorko, CS .
ANNALS OF PHARMACOTHERAPY, 1999, 33 (7-8) :775-780
[5]  
Cosmi B, 2000, HAEMATOLOGICA, V85, P826
[6]   Accuracy of a portable international normalization ratio monitor in outpatients receiving long-term oral anticoagulant therapy: Comparison with a laboratory reference standard using clinically relevant criteria for agreement [J].
Douketis, JD ;
Lane, A ;
Milne, J ;
Ginsberg, JS .
THROMBOSIS RESEARCH, 1998, 92 (01) :11-17
[7]   Clinical impact of interlaboratory variation in International Normalized Ratio determinations [J].
Finck, KM ;
Doetkott, C ;
Miller, DR .
AMERICAN JOURNAL OF HEALTH-SYSTEM PHARMACY, 2001, 58 (08) :684-688
[8]  
FYFE D, 1991, J AM COLL CARDIOL, V17, P1152
[9]   A comparison of point-of-gate instruments designed for monitoring oral anticoagulation with standard laboratory methods [J].
Gosselin, R ;
Owings, JT ;
White, RH ;
Hutchinson, R ;
Branch, J ;
Mahackian, K ;
Johnston, M ;
Larkin, EC .
THROMBOSIS AND HAEMOSTASIS, 2000, 83 (05) :698-703
[10]   Is the international normalised ratio (INR) reliable? A trial of comparative measurements in hospital laboratory and primary care settings [J].
Hobbs, FDR ;
Fitzmaurice, DA ;
Murray, ET ;
Holder, R ;
Rose, PE ;
Roper, JL .
JOURNAL OF CLINICAL PATHOLOGY, 1999, 52 (07) :494-497