摘要
Background: Strength of graft fixation is the weakest link in anterior cruciate ligament reconstruction. Hypothesis: There is no difference in initial fixation strength between different hamstring tendon graft femoral fixation devices. Study Design: Randomized experimental study. Methods: Each of six devices was used in the fixation of 10 quadrupled human semitendinosus-gracilis tendon grafts in tunnels drilled in porcine femora and tested 10 times with a single-cycle load-to-failure test at a rate of 50 mm/min and 10 times with a 1500-cycle loading test between 50 and 200 N at one cycle every 2 seconds. The specimens that survived the cyclic loading were subjected to a single-cycle load-to-failure test. Results: The Bone Mulch Screw (1112 N) was strongest in the single-cycle load-to-failure test, followed by EndoButton CL (1086 N), RigidFix (868 N), SmartScrew ACL (794 N), BioScrew (589 N), and RCI screw (546 N). It also showed the lowest residual displacement (2.2 mm) and was strongest in the single-cycle load-to-failure test after cyclic loading. Conclusions: The Bone Mulch Screw was superior to all other devices. Clinical Relevance: Caution may be warranted in employing aggressive rehabilitation after reconstruction with these devices. Preconditioning of the graft-implant complex before fixation is important. (C) 2003 American Orthopaedic Society for Sports Medicine.
引用
收藏
页码:174 / 181
页数:8
相关论文
共 30 条
[1]
In vitro testing protocols for the cruciate ligaments and ligament reconstructions [J].
Beynnon B.D. ;
Amis A.A. .
Knee Surgery, Sports Traumatology, Arthroscopy, 1998, 6 (Suppl 1) :S70-S76
[2]
The effect of screw insertion torque on tendons fixed with spiked washers [J].
Beynnon, BD ;
Meriam, CM ;
Ryder, SH ;
Fleming, BC ;
Johnson, RJ .
AMERICAN JOURNAL OF SPORTS MEDICINE, 1998, 26 (04) :536-539
[3]
Graft fixation in cruciate ligament reconstruction [J].
Brand, J ;
Weiler, A ;
Caborn, DNM ;
Brown, CH ;
Johnson, DL .
AMERICAN JOURNAL OF SPORTS MEDICINE, 2000, 28 (05) :761-774
[4]
Interference screw fixation strength of a quadrupled hamstring tendon graft is directly related to bone mineral density and insertion torque [J].
Brand, JC ;
Pienkowski, D ;
Steenlage, E ;
Hamilton, D ;
Johnson, DL ;
Caborn, DNM .
AMERICAN JOURNAL OF SPORTS MEDICINE, 2000, 28 (05) :705-710
[5]
Quadrupled semitendinosus-gracilis autograft fixation in the femoral tunnel: A comparison between a metal and a bioabsorbable interference screw [J].
Caborn, DNM ;
Coen, M ;
Neef, R ;
Hamilton, D ;
Nyland, J ;
Johnson, DL .
ARTHROSCOPY, 1998, 14 (03) :241-245
[6]
Cross-pin femoral fixation: A new technique for hamstring anterior cruciate ligament reconstruction of the knee [J].
Clark, R ;
Olsen, RE ;
Larson, BJ ;
Goble, EM ;
Farrer, RP .
ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 1998, 14 (03) :258-267
[7]
THE STRENGTH OF THE CENTRAL 3RD PATELLAR TENDON GRAFT - A BIOMECHANICAL STUDY [J].
COOPER, DE ;
DENG, XHH ;
BURSTEIN, AL ;
WARREN, RF .
AMERICAN JOURNAL OF SPORTS MEDICINE, 1993, 21 (06) :818-824
[8]
Current trends in anterior cruciate ligament reconstruction Part II. Operative procedures and clinical correlations [J].
Fu, FH ;
Bennett, CH ;
Ma, CB ;
Menetrey, J ;
Lattermann, C .
AMERICAN JOURNAL OF SPORTS MEDICINE, 2000, 28 (01) :124-130
[9]
Current trends in anterior cruciate ligament reconstruction part 1: Biology and biomechanics of reconstruction [J].
Fu, FH ;
Bennett, CH ;
Lattermann, C ;
Ma, CB .
AMERICAN JOURNAL OF SPORTS MEDICINE, 1999, 27 (06) :821-830
[10]
Comparative pull-out and cyclic-loading strength tests of anchorage of hamstring tendon grafts in anterior cruciate ligament reconstruction [J].
Giurea, M ;
Zorilla, P ;
Amis, AA ;
Aichroth, P .
AMERICAN JOURNAL OF SPORTS MEDICINE, 1999, 27 (05) :621-625