Cone-beam reconstruction using the backprojection of locally filtered projections

被引:175
作者
Pack, JD [1 ]
Noo, F [1 ]
Clackdoyle, R [1 ]
机构
[1] Univ Utah, Dept Radiol, Salt Lake City, UT 84108 USA
关键词
cone-beam tomography; helical CT; image reconstruction; saddle trajectories; truncated data;
D O I
10.1109/TMI.2004.837794
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
This paper describes a flexible new methodology for accurate cone beam reconstruction with source positions on a curve (or set of curves). The inversion formulas employed by this methodology are based on first backprojecting a simple derivative in the projection space and then applying a Hilbert transform inversion in the image space. The local nature of the projection space filtering distinguishes this approach from conventional filtered-backprojection methods. This characteristic together with a degree of flexibility in choosing the direction of the Hilbert transform used for inversion offers two important features for the design of data acquisition geometries and reconstruction algorithms. First, the size of the detector necessary to acquire sufficient data for accurate reconstruction of a given region is often smaller than that required by previously documented approaches. In other words, more data truncation is allowed. Second, redundant data can be incorporated for the purpose of noise reduction. The validity of the inversion formulas along with the application of these two properties are illustrated with reconstructions from computer simulated data. In particular, in the helical cone beam geometry, it is shown that 1) intermittent transaxial truncation has no effect on the reconstruction in a central region which means that wider patients can be accommodated on existing scanners, and more importantly that radiation exposure can be reduced for region of interest imaging and 2) at maximum pitch the data outside the Tam-Danielsson window can be used to reduce image noise and thereby improve dose utilization. Furthermore, the degree of axial truncation tolerated by our approach for saddle trajectories is shown to be larger than that of previous methods.
引用
收藏
页码:70 / 85
页数:16
相关论文
共 31 条
[1]   An alternative derivation of Katsevich's cone-beam reconstruction formula [J].
Chen, GH .
MEDICAL PHYSICS, 2003, 30 (12) :3217-3226
[2]  
CLACKDOYLE R, 1999, COMPUTATIONAL RADIOL, V110, P113
[3]  
DANIELSSON PE, 1997, P 1997 INT M FULL 3, P141
[4]   PRACTICAL CONE-BEAM ALGORITHM [J].
FELDKAMP, LA ;
DAVIS, LC ;
KRESS, JW .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06) :612-619
[5]  
FLOHR T, 2003, MED PHYS, V30
[6]  
GRANGEAT P, 1991, LECT NOTES MATH, V1497, P66
[7]   Redundant data and exact helical cone-beam reconstruction [J].
Heuscher, D ;
Brown, K ;
Noo, F .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (11) :2219-2238
[8]   Feldkamp and circle-and-line cone-beam reconstruction for 3D micro-CT of vascular networks [J].
Johnson, RH ;
Hu, H ;
Haworth, ST ;
Cho, PS ;
Dawson, CA ;
Linehan, JH .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (04) :929-940
[9]   An improved exact filtered backprojection algorithm for spiral computed tomography [J].
Katsevich, A .
ADVANCES IN APPLIED MATHEMATICS, 2004, 32 (04) :681-697
[10]  
Katsevich A, 2003, INT J MATH MATH SCI, V2003, DOI DOI 10.1155/S0161171203209315