Continuous estimation of systolic blood pressure using the pulse arrival time and intermittent calibration

被引:336
作者
Chen, W [1 ]
Kobayashi, T
Ichikawa, S
Takeuchi, Y
Togawa, T
机构
[1] Soka Univ, Fac Engn, Dept Bioengn, Tokyo, Japan
[2] Tokyo Womens Med Univ, Daini Hosp, Dept Cardiovasc Surg, Tokyo, Japan
[3] Tokyo Med & Dent Univ, Inst Biomat & Bioengn, Tokyo, Japan
关键词
photoplethysmogram; pulse arrival time; systolic blood pressure; noninvasive measurement;
D O I
10.1007/BF02345755
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
A continuous noninvasive method of systolic blood pressure estimation is described. Systolic blood pressure is estimated by combining two separately obtained components: a higher frequency component obtained by extracting a specific frequency band of pulse arrival time and a lower frequency component obtained from the intermittently acquired systolic blood pressure measurements with an auscultatory or oscillometric system. The pulse arrival time was determined by the time interval from ORS apex in electrocardiogram to the onset of photoplethysmogram in a fingertip beat-by-beat via an oximetric sensor. The method was examined in 20 patients during cardiovascular surgery. The estimated values of systolic blood pressure were compared with those measured invasively using a radial arterial catheter. The results showed that the correlation coefficients between estimated values and invasively obtained systolic blood pressure reached 0.97 +/- 0.02 (mean +/- SD), and the error remained within +/- 10% in 97.8% of the monitoring period. By using a system with automatic cuff inflation and deflation to acquire intermittent systolic blood pressure values, this method can be applicable for the continuous noninvasive monitoring of systolic blood pressure.
引用
收藏
页码:569 / 574
页数:6
相关论文
共 26 条
[1]   Measurements of pulse wave velocity [J].
Bazett, HC ;
Dreyer, NB .
AMERICAN JOURNAL OF PHYSIOLOGY, 1922, 63 (01) :94-116
[2]   Aortic pulse wave velocity as a marker of cardiovascular risk in hypertensive patients [J].
Blacher, J ;
Asmar, R ;
Djane, S ;
London, GM ;
Safar, ME .
HYPERTENSION, 1999, 33 (05) :1111-1117
[3]   RELATIONSHIP BETWEEN PULSE-WAVE VELOCITY AND ARTERIAL ELASTICITY [J].
CALLAGHAN, FJ ;
GEDDES, LA ;
BABBS, CF ;
BOURLAND, JD .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1986, 24 (03) :248-254
[4]   Study of arterial blood pressure by a Windkessel-type model:: influence of arterial functional properties [J].
Diourté, B ;
Siché, JP ;
Comparat, V ;
Baguet, JP ;
Mallion, JM .
COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 1999, 60 (01) :11-22
[5]   PULSE TRANSIT-TIME AS AN INDICATOR OF ARTERIAL BLOOD-PRESSURE [J].
GEDDES, LA ;
VOELZ, MH ;
BABBS, CF ;
BOURLAND, JD ;
TACKER, WA .
PSYCHOPHYSIOLOGY, 1981, 18 (01) :71-74
[6]   PULSE-WAVE VELOCITY AS A MEASURE OF BLOOD-PRESSURE CHANGE [J].
GRIBBIN, B ;
STEPTOE, A ;
SLEIGHT, P .
PSYCHOPHYSIOLOGY, 1976, 13 (01) :86-90
[7]   THE DETERMINATION OF THE PROPAGATION VELOCITY OF THE ARTERIAL PULSE WAVE [J].
HAMILTON, WF ;
REMINGTON, JW ;
DOW, P .
AMERICAN JOURNAL OF PHYSIOLOGY, 1945, 144 (04) :521-535
[8]   The extensibility of human arteries [J].
Hemingway, A ;
McSwiney, BA ;
Allison, PR .
QUARTERLY JOURNAL OF MEDICINE, 1928, 21 (84) :489-498
[9]  
HICKSON SK, 1924, J PHYSL, V19, P217
[10]  
HON EH, 1992, OBSTET GYNECOL, V79, P597