Comparison of BACTEC MGIT 960 and BACTEC 460 for culture of Mycobacteria

被引:13
作者
Whyte, T
Cormican, M [1 ]
Hanahoe, B
Doran, G
Collins, T
Corbett-Feeney, G
机构
[1] Univ Coll Hosp Galway, Dept Med Microbiol, Galway, Ireland
[2] Natl Univ Ireland, Dept Bacteriol, Galway, Ireland
关键词
Mycobacterium; culture; Bactec MGIT; Bactec; 460;
D O I
10.1016/S0732-8893(00)00179-6
中图分类号
R51 [传染病];
学科分类号
100401 ;
摘要
We have compared the BACTEC 460 system with the BACTEC MGIT 960 system for culture of mycobacteria from 1800 routine clinical specimens. Rate of isolation of M. tuberculosis and time to detection of positive culture was comparable for both systems (BACTEC 460, 35 isolates, BACTEC MGIT 960, 34 isolates). Contamination of cultures was more common with the BACTEC MGIT 960 system. With intensification of the decontamination process an acceptable contamination rate was achieved in the BACTEC MGIT 960 system but time to detection of positive culture was increased by 1 to 2 days. (C) 2000 Elsevier Science Inc. All rights reserved.
引用
收藏
页码:123 / 126
页数:4
相关论文
共 12 条
[11]   Turnaround times for mycobacterial cultures [J].
Styrt, BA ;
Shinnick, TM ;
Ridderhof, JC ;
Crawford, JT ;
Tenover, FC .
JOURNAL OF CLINICAL MICROBIOLOGY, 1997, 35 (04) :1041-1041
[12]   Use of BACTEC MGIT 960 for recovery of mycobacteria from clinical specimens: Multicenter study [J].
Tortoli, E ;
Cichero, P ;
Piersimoni, C ;
Simonetti, MT ;
Gesu, G ;
Nista, D .
JOURNAL OF CLINICAL MICROBIOLOGY, 1999, 37 (11) :3578-3582