A numerical approach to the determination of mass transfer performances through partially wetted microporous membranes: transfer of oxygen to water

被引:26
作者
Evren, V
机构
[1] Hacettepe University, Chem. Eng. Dept., Beytepe, 06532, Ankara
关键词
membrane oxygenators; microporous membranes; mass transfer modeling; numerical solution; wetted pores;
D O I
10.1016/S0376-7388(00)00401-4
中图分类号
TQ [化学工业];
学科分类号
0817 ;
摘要
Microporous polypropylene membranes were used to transfer oxygen to water where the gas phase was pure oxygen saturated with water vapor The experiments were maintained at a constant temperature of 37 degrees C. Flow channel length, channel thickness, liquid velocity and operating pressure were changed so as to investigate their effects on mass transfer performances. A model for non-wetted pores was derived, by neglecting the gas phase film and pore resistances, and numerically solved in order to find the concentration distribution and the local average concentration in the liquid phase. This model was further developed for wetted ports in order to reflect the true operating conditions. Wetting ratio was found by the matching of model predicted concentration profiles with the experimental results. Oxygen flux and liquid side film coefficient were also calculated by using experimental data. All of the results of this work were in agreement with the literature. (C) 2000 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:97 / 110
页数:14
相关论文
共 18 条
[1]   USE OF SEALED END HOLLOW FIBERS FOR BUBBLELESS MEMBRANE AERATION - EXPERIMENTAL STUDIES [J].
AHMED, T ;
SEMMENS, MJ .
JOURNAL OF MEMBRANE SCIENCE, 1992, 69 (1-2) :1-10
[2]   SOLVENT-EXTRACTION OF PHENOL FROM AQUEOUS-SOLUTION IN A HOLLOW FIBER DEVICE [J].
COONEY, DO ;
JIN, CL .
CHEMICAL ENGINEERING COMMUNICATIONS, 1985, 37 (1-6) :173-191
[3]  
DORSON W, 1968, T AM SOC ART INT ORG, V14, P242
[4]  
EBERHART RC, 1978, ARTIF ORGANS, V2, P19
[5]  
EVREN V, 1988, THESIS HACETTEPE U A
[6]  
GEANKOPLIS CJ, 1993, TRANSPORT PROCESSES, pCH5
[7]   Characterization of microporous membranes for use in membrane contactors [J].
Iversen, SB ;
Bhatia, VK ;
DamJohansen, K ;
Jonsson, G .
JOURNAL OF MEMBRANE SCIENCE, 1997, 130 (1-2) :205-217
[8]   GAS-ABSORPTION STUDIES IN MICROPOROUS HOLLOW FIBER MEMBRANE MODULES [J].
KAROOR, S ;
SIRKAR, KK .
INDUSTRIAL & ENGINEERING CHEMISTRY RESEARCH, 1993, 32 (04) :674-684
[9]   DETERMINATION OF MASS-TRANSFER RATES IN WETTED AND NON-WETTED MICROPOROUS MEMBRANES [J].
KREULEN, H ;
SMOLDERS, CA ;
VERSTEEG, GF ;
VANSWAAIJ, WPM .
CHEMICAL ENGINEERING SCIENCE, 1993, 48 (11) :2093-2102
[10]   LOW-ORDER APPROXIMATIONS FOR MEMBRANE BLOOD OXYGENATORS [J].
LIGHTFOOT, EN .
AICHE JOURNAL, 1968, 14 (04) :669-+