Implantation of retina stimulation electrodes and recording of electrical stimulation responses in the visual cortex of the cat

被引:79
作者
Hesse, L
Schanze, T
Wilms, M
Eger, M
机构
[1] Univ Marburg, Dept Ophthalmol, D-35033 Marburg, Germany
[2] Univ Marburg, Dept Neurophys, Marburg, Germany
关键词
D O I
10.1007/s004170000197
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background: Simple basic visual perception may be restored by epiretinal electrical stimulation in patients that are blind due to photoreceptor loss. To stimulate ganglion cells, epiretinally flat platinum microelectrodes embedded in thin polyimide film were developed and tested in the cat. Method's: After removal of the lens and the vitreous body a thin microfilm electrode array was implanted through a corneoscleral incision in the cat eye (n=4). In two eyes no further attempt was made to fixate the tip of the electrode, which was pressed onto the retinal surface due to the tension of the curved polyimide film. In two eyes the tip of the electrode was fixed with cyanoacrylate adhesive. The exterior part of the microelectrode film was directed under the skin towards the forehead which allowed fixation of the microplug to a head fixation bolt. Retinal stimulation experiments were performed within 1 week after implantation. Success of stimulation was assessed by recording neuronal activities from areas 17 and 18. Retinal microelectrodes were removed 7 weeks or longer after implantation. Results: Intraocular inflammation or retinal detachment were not observed after implantation of the microelectrode film. In two eyes the tip of the microelectrodes dislocated spontaneously within the first few days. The lowest threshold of electrical stimulation was 35 muA, corresponding to a charge transfer of 14 nC per phase. These values were ten times higher than those obtained by needle electrodes used in prior experiments. Conclusions: Intraocular implanted fiat microelectrodes made of platinum and polyimide were well tolerated. Because of the flat configuration of the microelectrodes higher stimulation thresholds than for needle electrodes were found, indicating insufficient contact to the retinal surface. An alternative shape and fixation technique is required res minimise electrodes' threshold of stimulation.
引用
收藏
页码:840 / 845
页数:6
相关论文
共 18 条
[1]   ELECTRICAL-STIMULATION WITH PT ELECTRODES .2. ESTIMATION OF MAXIMUM SURFACE REDOX (THEORETICAL NON-GASSING) LIMITS [J].
BRUMMER, SB ;
TURNER, MJ .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1977, 24 (05) :440-443
[2]   Learning retina implants with epiretinal contacts [J].
Eckmiller, R .
OPHTHALMIC RESEARCH, 1997, 29 (05) :281-289
[3]   Cyanoacrylate glue in the repair of retinal detachment associated with posterior retinal breaks in infants and children [J].
Hartnett, ME ;
Hirose, T .
RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 1998, 18 (02) :125-129
[4]   Induction of posterior vitreous detachment in rabbits by intravitreal injection of tissue plasminogen activator following cryopexy [J].
Hesse, L ;
Nebeling, B ;
Schroeder, B ;
Heller, G ;
Kroll, P .
EXPERIMENTAL EYE RESEARCH, 2000, 70 (01) :31-39
[5]   RETINAL TOXICITY OF CYANOACRYLATE TISSUE ADHESIVE IN THE RABBIT [J].
HIDA, T ;
SHETA, SM ;
PROIA, AD ;
MCCUEN, BW .
RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 1988, 8 (02) :148-153
[6]   Pattern electrical stimulation of the human retina [J].
Humayun, MS ;
de Juan, E ;
Weiland, JD ;
Dagnelie, G ;
Katona, S ;
Greenberg, R ;
Suzuki, S .
VISION RESEARCH, 1999, 39 (15) :2569-2576
[7]   Visual perception elicited by electrical stimulation of retina in blind humans [J].
Humayun, MS ;
deJuan, E ;
Dagnelie, G ;
Greenberg, RJ ;
Prost, RH ;
Phillips, DH .
ARCHIVES OF OPHTHALMOLOGY, 1996, 114 (01) :40-46
[8]  
Humayun MS, 1999, INVEST OPHTH VIS SCI, V40, P143
[9]  
JOHNSON MW, 1992, OPHTHALMOLOGY, V99, P515
[10]  
Majji AB, 1999, INVEST OPHTH VIS SCI, V40, P2073