High-speed peak matching algorithm for retention time alignment of gas chromatographic data for chemometric analysis

被引:152
作者
Johnson, KJ
Wright, BW
Jarman, KH
Synovec, RE
机构
[1] Univ Washington, Dept Chem, Ctr Proc Analyt Chem, Seattle, WA 98195 USA
[2] Pacific NW Natl Lab, Richland, WA 99352 USA
关键词
peak matching algorithm; retention times; diesel fuel; principal component analysis;
D O I
10.1016/S0021-9673(03)00616-2
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A rapid retention time alignment algorithm was developed as a preprocessing utility to be used prior to chemometric analysis of large datasets of diesel fuel profiles obtained using gas chromatography (GC). Retention time variation from chromatogram-to-chromatogram has been a significant impediment against the use of chemometric techniques in the analysis of chromatographic data due to the inability of current chemometric techniques to correctly model information that shifts from variable to variable within a dataset. The alignment algorithm developed is shown to increase the efficacy of pattern recognition methods applied to diesel fuel chromatograms by retaining chemical selectivity while reducing chromatogram-to-chromatogram retention time variations and to do so on a time scale that makes analysis of large sets of chromatographic data practical. Two sets of diesel fuel gas chromatograms were studied using the novel alignment algorithm followed by principal component analysis (PCA). In the first study, retention times for corresponding chromatographic peaks in 60 chromatograms varied by as much as 300 ms between chromatograms before alignment. In the second study of 42 chromatograms, the retention time shifting exhibited was on the order of 10 s between corresponding chromatographic peaks, and required a coarse retention time correction prior to alignment with the algorithm. In both cases, an increase in retention time precision afforded by the algorithm was clearly visible in plots of overlaid chromatograms before and then after applying the retention time alignment algorithm. Using the alignment algorithm, the standard deviation for corresponding peak retention times following alignment was 17 ms throughout a given chromatogram, corresponding to a relative standard deviation of 0.003% at an average retention time of 8 min. This level of retention time precision is a 5-fold improvement over the retention time precision initially provided by a state-of-the-art GC instrument equipped with electronic pressure control and was critical to the performance of the chemometric analysis. This increase in retention time precision does not come at the expense of chemical selectivity, since the PCA results suggest that essentially all of the chemical selectivity is preserved. Cluster resolution between dissimilar groups of diesel fuel chromatograms in a two-dimensional scores space generated with PCA is shown to substantially increase after alignment. The alignment method is robust against missing or extra peaks relative to a target chromatogram used in the alignment, and operates at high speed, requiring roughly 1 s of computation time per GC chromatogram. (C) 2003 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:141 / 155
页数:15
相关论文
共 9 条
[1]   SIMPLEX FOCUSING OF RETENTION TIMES AND LATENT VARIABLE PROJECTIONS OF CHROMATOGRAPHIC PROFILES [J].
ANDERSSON, R ;
HAMALAINEN, MD .
CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, 1994, 22 (01) :49-61
[2]   CORRELATION OF QUANTITATIVE-ANALYSIS PRECISION TO RETENTION TIME PRECISION AND CHROMATOGRAPHIC RESOLUTION FOR RAPID, SHORT COLUMN ANALYSIS [J].
BAHOWICK, TJ ;
SYNOVEC, RE .
ANALYTICAL CHEMISTRY, 1995, 67 (03) :631-640
[3]   Increasing the number of analyzable peaks in comprehensive two dimensional separations through chemometrics [J].
Fraga, CG ;
Bruckner, CA ;
Synovec, RE .
ANALYTICAL CHEMISTRY, 2001, 73 (03) :675-683
[4]  
Malinowski E.R., 1991, FACTOR ANAL CHEM
[5]   ALIGNMENT OF CHROMATOGRAPHIC PROFILES FOR PRINCIPAL COMPONENT ANALYSIS - A PREREQUISITE FOR FINGERPRINTING METHODS [J].
MALMQUIST, G ;
DANIELSSON, R .
JOURNAL OF CHROMATOGRAPHY A, 1994, 687 (01) :71-88
[6]  
MAYFIELD HT, 1983, COMPUT APPL LAB, V1, P130
[7]   Aligning of single and multiple wavelength chromatographic profiles for chemometric data analysis using correlation optimised warping [J].
Nielsen, NPV ;
Carstensen, JM ;
Smedsgaard, J .
JOURNAL OF CHROMATOGRAPHY A, 1998, 805 (1-2) :17-35
[8]   EFFECTS OF RESOLUTION, PEAK RATIO AND SAMPLING FREQUENCY IN DIODE-ARRAY FLUORESCENCE DETECTION IN LIQUID-CHROMATOGRAPHY [J].
POE, RB ;
RUTAN, SC .
ANALYTICA CHIMICA ACTA, 1993, 283 (02) :845-853
[9]   Standardization of second-order chromatographic/spectroscopic data for optimum chemical analysis [J].
Prazen, BJ ;
Synovec, RE ;
Kowalski, BR .
ANALYTICAL CHEMISTRY, 1998, 70 (02) :218-225