Multicenter study of whole-blood creatinine, total carbon dioxide content, and chemistry profiling for laboratory and point-of-care testing in critical care in the United States

被引:23
作者
Kost, GJ
Vu, HT
Inn, M
DuPlantier, R
Fleisher, M
Kroll, MH
Spinosa, JC
机构
[1] Univ Calif Davis, Sch Med, POCT CTR, Davis, CA 95616 USA
[2] Mem Sloan Kettering Canc Ctr, New York, NY 10021 USA
[3] Scripps Mem Hosp, La Jolla, CA USA
关键词
biosensor; carbon dioxide; chloride; emergency room; glucose; hematocrit; operating room; potassium; precision; sodium; substrate-specific electrode; urea nitrogen; and whole-blood analyzer;
D O I
10.1097/00003246-200007000-00033
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objectives: To introduce a creatinine biosensor and a total carbon dioxide content (TCO2) method for whole-blood measurements, to evaluate the clinical performance of a new transportable analyzer that simultaneously performs these two and six other tests (Na+, K+, Cl-, glucose, urea nitrogen, and hematocrit), and to assess the potential of the new analyzer for point-of-care testing in critical care by comparing results obtained by nonlaboratory personnel and by medical technologists. Design: Multicenter sites compared whole-blood measurements with the transportable analyzer to plasma measurements from the same specimens with local reference instruments. One site compared whole-blood results produced by nonlaboratory personnel vs. medical technologists and evaluated day-to-day and within-day precision at the point of care. Settings and Patients: Four medical centers in the United States. Venous and arterial specimens from 710 critically ill patients with a variety of diagnoses. Point-of-care testing in the emergency room and operating room. Results: The linear regression analyses at the four medical centers showed the following: creatinine (a) slope, 0.91 to 1.22, (b) y intercept, -0.07 to 0.15 mg/dL, and (c) r(2), 0.77 to 1.00; and TCO2: (a) slope, 0.64 to 1.00, (b) y intercept, 1.36 to 9.6 mmol/L, and (c) r(2), 0.52 to 0.72 (y(i), whole-blood analyses; x(i) plasma reference measurements). Bland-Altman plots also were used to assess multicenter creatinine and TCO2 results. Of the other analytes, K+, glucose, and urea nitrogen had the highest r(2)-values. For the eight chemistry profile tests performed at the point of care (y(i), nonlaboratory personnel results; x(i) medical technologist results), the average value of r(2) was 0.96 (SD 0.08) in the operating room and 0.96 (SD 0.06) in the emergency room, and mean paired differences (y(i) - x(i)) were not statistically or clinically significant. Precision was acceptable. Conclusions: The performance of the creatinine biosensor and the TCO2 method was acceptable for whole-blood samples. Comparisons of whole-blood results from the transportable analyzer and plasma results from the local reference instruments revealed analyte biases that may be attributed to differences between direct whole-blood analyses and indirect-diluted plasma measurements and other factors. Performance of nonlaboratory personnel and medical technologists was equivalent for point-of-care testing in critical care settings. The whole-blood analyzer should be useful when patient care demands immediate results.
引用
收藏
页码:2379 / 2389
页数:11
相关论文
共 46 条
[1]  
[Anonymous], 1992, FED REG
[2]  
[Anonymous], 1998, PRINCIPLES PRACTICE
[3]  
APPLE FS, 1982, CLIN CHEM, V28, P1931
[4]  
BISHOP MS, 1994, ARCH PATHOL LAB MED, V118, P797
[5]  
BLIJENBERG BG, 1994, EUR J CLIN CHEM CLIN, V32, P529
[6]   The magnitude of metabolic acidosis is dependent on differences in bicarbonate assays [J].
Bray, SH ;
Tung, RL ;
Jones, ER .
AMERICAN JOURNAL OF KIDNEY DISEASES, 1996, 28 (05) :700-703
[7]   Measurement of blood CO2 concentration with a conventional Pco(2) analyzer [J].
Breen, PH ;
Mazumdar, B ;
Skinner, SC ;
Taitelman, UZ ;
Isserles, SA .
CRITICAL CARE MEDICINE, 1996, 24 (07) :1215-1218
[8]  
COWELL DC, 1985, CLIN CHEM, V31, P2009
[9]  
DOUGLAS AR, 1988, J APPL PHYSIOL, V65, P473
[10]  
EHRMEYER SS, 1995, AM J CLIN PATHOL, V104, pS40