Multicenter study of whole-blood creatinine, total carbon dioxide content, and chemistry profiling for laboratory and point-of-care testing in critical care in the United States

被引:23
作者
Kost, GJ
Vu, HT
Inn, M
DuPlantier, R
Fleisher, M
Kroll, MH
Spinosa, JC
机构
[1] Univ Calif Davis, Sch Med, POCT CTR, Davis, CA 95616 USA
[2] Mem Sloan Kettering Canc Ctr, New York, NY 10021 USA
[3] Scripps Mem Hosp, La Jolla, CA USA
关键词
biosensor; carbon dioxide; chloride; emergency room; glucose; hematocrit; operating room; potassium; precision; sodium; substrate-specific electrode; urea nitrogen; and whole-blood analyzer;
D O I
10.1097/00003246-200007000-00033
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objectives: To introduce a creatinine biosensor and a total carbon dioxide content (TCO2) method for whole-blood measurements, to evaluate the clinical performance of a new transportable analyzer that simultaneously performs these two and six other tests (Na+, K+, Cl-, glucose, urea nitrogen, and hematocrit), and to assess the potential of the new analyzer for point-of-care testing in critical care by comparing results obtained by nonlaboratory personnel and by medical technologists. Design: Multicenter sites compared whole-blood measurements with the transportable analyzer to plasma measurements from the same specimens with local reference instruments. One site compared whole-blood results produced by nonlaboratory personnel vs. medical technologists and evaluated day-to-day and within-day precision at the point of care. Settings and Patients: Four medical centers in the United States. Venous and arterial specimens from 710 critically ill patients with a variety of diagnoses. Point-of-care testing in the emergency room and operating room. Results: The linear regression analyses at the four medical centers showed the following: creatinine (a) slope, 0.91 to 1.22, (b) y intercept, -0.07 to 0.15 mg/dL, and (c) r(2), 0.77 to 1.00; and TCO2: (a) slope, 0.64 to 1.00, (b) y intercept, 1.36 to 9.6 mmol/L, and (c) r(2), 0.52 to 0.72 (y(i), whole-blood analyses; x(i) plasma reference measurements). Bland-Altman plots also were used to assess multicenter creatinine and TCO2 results. Of the other analytes, K+, glucose, and urea nitrogen had the highest r(2)-values. For the eight chemistry profile tests performed at the point of care (y(i), nonlaboratory personnel results; x(i) medical technologist results), the average value of r(2) was 0.96 (SD 0.08) in the operating room and 0.96 (SD 0.06) in the emergency room, and mean paired differences (y(i) - x(i)) were not statistically or clinically significant. Precision was acceptable. Conclusions: The performance of the creatinine biosensor and the TCO2 method was acceptable for whole-blood samples. Comparisons of whole-blood results from the transportable analyzer and plasma results from the local reference instruments revealed analyte biases that may be attributed to differences between direct whole-blood analyses and indirect-diluted plasma measurements and other factors. Performance of nonlaboratory personnel and medical technologists was equivalent for point-of-care testing in critical care settings. The whole-blood analyzer should be useful when patient care demands immediate results.
引用
收藏
页码:2379 / 2389
页数:11
相关论文
共 46 条
[21]  
Kost G J, 1991, J Int Fed Clin Chem, V3, P160
[22]  
Kost G J, 1992, MLO Med Lab Obs, V24, P17
[23]  
KOST GJ, 1995, AM J CLIN PATHOL, V104, pS2
[24]  
KOST GJ, 1988, CLIN CHEM, V34, P1650
[25]  
KOST GJ, 1990, ARCH PATHOL LAB MED, V114, P865
[26]   NEW WHOLE-BLOOD ANALYZERS AND THEIR IMPACT ON CARDIAC AND CRITICAL CARE [J].
KOST, GJ .
CRITICAL REVIEWS IN CLINICAL LABORATORY SCIENCES, 1993, 30 (02) :153-202
[27]   MONITORING OF IONIZED CALCIUM DURING HUMAN HEPATIC TRANSPLANTATION - CRITICAL-VALUES AND THEIR RELEVANCE TO CARDIAC AND HEMODYNAMIC MANAGEMENT [J].
KOST, GJ ;
JAMMAL, MA ;
WARD, RE ;
SAFWAT, AM .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1986, 86 (01) :61-70
[28]  
KOST GJ, 1995, AM J CLIN PATHOL, V104, pS111
[29]   Multicenter study of oxygen-insensitive handheld glucose point-of-care testing in critical care hospital ambulatory patients in the United States and Canada [J].
Kost, GJ ;
Vu, HT ;
Lee, JH ;
Bourgeois, P ;
Kiechle, FL ;
Martin, C ;
Miller, SS ;
Okorodudu, AO ;
Podczasy, JJ ;
Webster, R ;
Whitlow, KJ .
CRITICAL CARE MEDICINE, 1998, 26 (03) :581-590
[30]  
KOST GJ, 1996, HDB CLIN AUTOMATION, P757