摘要
This work presents a novel flow switch for centrifugal microfluidic platforms which is solely controlled by the Coriolis pseudo force. This Coriolis switch consists of an inverse Y-structure with one common upstream channel and two symmetric outlets on a rotating disk. Above a certain threshold frequency, the Coriolis force becomes dominant that the entire flow is diverted into one of the outlets which is selected by the direction of rotation. The threshold frequency has been measured to be 350 rad s(-1) (similar to55.7 Hz) for a channel width of 360 mm and a depth of 125 mum. The results are supported by extensive CFD simulations.
引用
收藏
页码:146 / 150
页数:5
相关论文
共 12 条
  • [1] BRENNER T, 2003, MESA MONOGRAPHS, P603
  • [2] BRENNER T, 2003, P 2 VDE WORLD MICR C, P171
  • [3] DUCREE J, 2003, MESA MONOGRAPHS, P903
  • [4] DUCREE J, BIODISK CENTRIFUGAL
  • [5] Microfabricated centrifugal microfluidic systems: Characterization and multiple enzymatic assays
    Duffy, DC
    Gillis, HL
    Lin, J
    Sheppard, NF
    Kellogg, GJ
    [J]. ANALYTICAL CHEMISTRY, 1999, 71 (20) : 4669 - 4678
  • [6] Ekstrand G, 2000, MESA MG, P311
  • [7] INGANAS M, 2001, P MICR TOT AN SYST, P91
  • [8] Madou M. J., 2001, SENSOR ACTUAT A-PHYS, V91, P301
  • [9] Design and Fabrication of CD-like Microfluidic Platforms for Diagnostics: Microfluidic Functions
    Madou, Marc J.
    Lee, L. James
    Daunert, Sylvia
    Lai, Siyi
    Shih, Chih-Hsin
    [J]. BIOMEDICAL MICRODEVICES, 2001, 3 (03) : 245 - 254
  • [10] The LabCD™:: A centrifuge-based microfluidic platform for diagnostics
    Madou, MJ
    Kellogg, GJ
    [J]. SYSTEMS AND TECHNOLOGIES FOR CLINICAL DIAGNOSTICS AND DRUG DISCOVERY, PROCEEDINGS OF, 1998, 3259 : 80 - 93