Determination of digitised radiograph magnification factors for pre-operative templating in hip prosthesis surgery

被引:16
作者
Descamps, Stephane [1 ]
Livesey, Christine [2 ]
Learmonth, Ian Douglas [2 ]
机构
[1] CHU Clermont Ferrand, Hop Gabriel Montpied, Serv Chirurg Orthoped & Traumatol, F-63003 Clermont Ferrand 01, France
[2] Southmead Hosp, Avon Orthopaed Ctr, Bristol Implant Res Ctr, Bristol BS10 5NB, Avon, England
关键词
Digital radiograph; Template; Magnification factor; Hip surgery; PELVIC X-RAYS; DIGITAL RADIOGRAPHS; ARTHROPLASTY; ACCURACY; REPRODUCIBILITY;
D O I
10.1007/s00256-009-0732-8
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
100224 [整形外科学];
摘要
With digital radiography development, information technology (IT) companies have developed specific software for templating procedures, requiring individual magnification assessments for each patient. The aim of this study was to determine the mean magnification factor of digital radiographs and to evaluate the possibility of using the mean magnification factor or clinical information in templating. We retrospectively evaluated 100 primary total hip arthroplasty digital radiographs using the femoral head prosthesis as a calliper to determinate the mean magnification factor. Working on the assumption that altitude of the hip during radiograph is decisive in modification of magnification factors, we also looked for a correlation between weight, body mass index (BMI), altitude and magnification factor. The magnification factor was 126% (121-130%). A relationship was found between magnification factor (Mf) and weight (Mf = 7.10(-4)x weight (kg) + 1.21), but not BMI. In 98% of cases, if the weight-correlated formula is used, the sizing is correct or the error is +/- 1 mm. With the mean method the sizing is correct or within 1 mm in only 78.2% of cases. Levels of accuracy for the mean magnification factor and the weight-correlated formula are not as high as individual assessments using a calliper; however, they could be used in everyday practice where individual magnification factors have not been calculated.
引用
收藏
页码:273 / 277
页数:5
相关论文
共 12 条
[1]
Digital templating in total hip arthroplasty [J].
Bono, JV .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A :118-122
[2]
Digital image analysis: improving accuracy and reproducibility of radiographic measurement [J].
Bould, M ;
Barnard, S ;
Learmonth, ID ;
Cunningham, JL ;
Hardy, JRW .
CLINICAL BIOMECHANICS, 1999, 14 (06) :434-437
[3]
A simple guide to determine the magnification of radiographs and to improve the accuracy of preoperative templating [J].
Conn, KS ;
Clarke, MT ;
Hallett, JP .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2002, 84B (02) :269-272
[4]
Surgical planning of total hip arthroplasty: accuracy of computer-assisted EndoMap software in predicting component size [J].
Davila, JA ;
Kransdorf, MJ ;
Duffy, GP .
SKELETAL RADIOLOGY, 2006, 35 (06) :390-393
[5]
Preoperative planning for primary total hip arthroplasty [J].
Della Valle, AG ;
Padgett, DE ;
Salvati, EA .
JOURNAL OF THE AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS, 2005, 13 (07) :455-462
[6]
Knight J L, 1992, J Arthroplasty, V7 Suppl, P403, DOI 10.1016/S0883-5403(07)80031-3
[7]
Linclau L, 1993, Acta Orthop Belg, V59, P163
[8]
Assessment of reproducibility and accuracy in templating hybrid total hip arthroplasty using digital radiographs [J].
Oddy, MJ ;
Jones, MJ ;
Pendegrass, CJ ;
Pilling, JR ;
Wimhurst, JA .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2006, 88B (05) :581-585
[9]
Digital correction of magnification in pelvic x rays for preoperative planning of hip joint replacements: Theoretical development and clinical results of a new protocol [J].
The, B ;
Diercks, RL ;
Stewart, RE ;
van Ooijen, PMA ;
van Horn, JR .
MEDICAL PHYSICS, 2005, 32 (08) :2580-2589
[10]
Comparison of analog and digital preoperative planning in total hip and knee arthroplasties - A prospective study of 173 hips and 65 total knees [J].
The, B ;
Diercks, RL ;
van Ooijen, PMA ;
van Horn, JR .
ACTA ORTHOPAEDICA, 2005, 76 (01) :78-84