Photoplethysmography and its application in clinical physiological measurement

被引:2496
作者
Allen, John [1 ]
机构
[1] Freeman Rd Hosp, Dept Reg Med Phys, Newcastle Upon Tyne NE7 7DN, Tyne & Wear, England
关键词
ageing; artery; autonomic function; blood pressure; cardiac output; cardiovascular; diabetes; endothelial function; heart rate; infrared; microcirculation; photoplethysmography (PPG); pulse wave analysis; Raynaud's phenomenon; vascular disease; vein; PULSE TRANSIT-TIME; PERIPHERAL VASCULAR-DISEASE; SYSTOLIC BLOOD-PRESSURE; RECORDING ANALYTICAL METHOD; AGE-RELATED-CHANGES; PHOTOELECTRIC PLETHYSMOGRAPHY; CARDIAC-OUTPUT; NONINVASIVE MEASUREMENT; NEURAL-NETWORK; WAVE VELOCITY;
D O I
10.1088/0967-3334/28/3/R01
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Photoplethysmography (PPG) is a simple and low-cost optical technique that can be used to detect blood volume changes in the microvascular bed of tissue. It is often used non-invasively to make measurements at the skin surface. The PPG waveform comprises a pulsatile ('AC') physiological waveform attributed to cardiac synchronous changes in the blood volume with each heart beat, and is superimposed on a slowly varying ('DC') baseline with various lower frequency components attributed to respiration, sympathetic nervous system activity and thermoregulation. Although the origins of the components of the PPG signal are not fully understood, it is generally accepted that they can provide valuable information about the cardiovascular system. There has been a resurgence of interest in the technique in recent years, driven by the demand for low cost, simple and portable technology for the primary care and community based clinical settings, the wide availability of low cost and small semiconductor components, and the advancement of computer-based pulse wave analysis techniques. The PPG technology has been used in a wide range of commercially available medical devices for measuring oxygen saturation, blood pressure and cardiac output, assessing autonomic function and also detecting peripheral vascular disease. The introductory sections of the topical review describe the basic principle of operation and interaction of light with tissue, early and recent history of PPG, instrumentation, measurement protocol, and pulse wave analysis. The review then focuses on the applications of PPG in clinical physiological measurements, including clinical physiological monitoring, vascular assessment and autonomic function.
引用
收藏
页码:R1 / R39
页数:39
相关论文
共 269 条
[61]   EFFECT OF THE ANATOMICAL STRUCTURE OF THE ARTERIAL TREE ON THE MEASUREMENT OF PULSE-WAVE VELOCITY IN MAN [J].
DELLACORTE, M ;
LOCCHI, F ;
SPINELLI, E ;
SCARPELLI, PT .
PHYSICS IN MEDICINE AND BIOLOGY, 1979, 24 (03) :593-599
[62]   WHAT DOES PHOTOPLETHYSMOGRAPHY MEASURE [J].
DETRAFFORD, J ;
LAFFERTY, K .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1984, 22 (05) :479-480
[63]   MODELING OF THE HUMAN VASOMOTOR CONTROL-SYSTEM AND ITS APPLICATION TO THE INVESTIGATION OF ARTERIAL-DISEASE [J].
DETRAFFORD, JC ;
LAFFERTY, K ;
KITNEY, RI ;
COTTON, LT ;
ROBERTS, VC .
IEE PROCEEDINGS-A-SCIENCE MEASUREMENT AND TECHNOLOGY, 1982, 129 (09) :646-650
[64]   The form of the volume pulse in the finger pad in health, arteriosclerosis, and hypertension [J].
Dillon, JB ;
Hertzman, AB .
AMERICAN HEART JOURNAL, 1941, 21 :172-190
[65]   Relation between heart rate and pulse transit time during paced respiration [J].
Drinnan, MJ ;
Allen, J ;
Murray, A .
PHYSIOLOGICAL MEASUREMENT, 2001, 22 (03) :425-432
[66]  
Duck F.A., 1990, Physical Properties of Tissue: A Comprehensive Reference Book
[67]  
DUTCH J, 1983, NEW ZEAL MED J, V96, P607
[68]   Vibration syndrome diagnosis using a cooling test verified by computerized photoplethysmography [J].
Dyszkiewicz, A ;
Tendera, M .
PHYSIOLOGICAL MEASUREMENT, 2006, 27 (04) :353-369
[69]  
ECHIADIS A, 2005, 314 LOUGHB U DEP EL
[70]   PULSE WAVE VELOCITY IN HEALTHY SUBJECTS AND IN PATIENTS WITH VARIOUS DISEASE STATES [J].
ELIAKIM, M ;
SAPOZNIKOV, D ;
WEINMAN, J .
AMERICAN HEART JOURNAL, 1971, 82 (04) :448-+