A low power, microvalve regulated architecture for drug delivery systems

被引:22
作者
Evans, Allan Thomas [1 ]
Park, Jong M. [1 ]
Chiravuri, Srinivas [2 ]
Gianchandani, Yogesh B. [1 ]
机构
[1] Univ Michigan, Dept Elect Engn & Comp Sci, Ann Arbor, MI 48109 USA
[2] Univ Michigan, Dept Anesthesiol, Ann Arbor, MI 48109 USA
关键词
Drug delivery; Intrathecal; Microvalve; Piezoelectric; Flow control; MEMS;
D O I
10.1007/s10544-009-9372-y
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper describes an actively-controlled architecture for drug delivery systems that offers high performance and volume efficiency through the use of micromachined components. The system uses a controlled valve to regulate dosing by throttling flow from a mechanically pressurized reservoir, thereby eliminating the need for a pump. To this end, the valve is fabricated from a glass wafer and silicon-on-insulator wafer for sensor integration. The valve draws a maximum power of 1.68 mu W (averaged over time); with the existing packaging scheme, it has a volume of 2.475 cm(3). The reservoirs are assembled by compressing polyethylene terephthalate polymer balloons with metal springs. The metal springs are fabricated from Elgiloy (R) using photochemical etching. The springs pressurize the contents of 37 mL chambers up to 15 kPa. The system is integrated with batteries and a control circuit board within a 113 cm(3) metal casing. This system has been evaluated in different control modes to mimic clinical applications. Bolus deliveries of 1.5 mL have been regulated as well as continuous flows of 0.15 mL/day with accuracies of 3.22%. The results suggest that this device can be used in an implant to regulate intrathecal drug delivery.
引用
收藏
页码:159 / 168
页数:10
相关论文
共 35 条
[1]  
*AM SOC HLTH SYST, 2000, AM J HEALTH-SYST PH, V57, P12
[2]   A prospective study of long-term intrathecal morphine in the management of chronic nonmalignant pain [J].
Anderson, VC ;
Burchiel, KJ .
NEUROSURGERY, 1999, 44 (02) :289-300
[3]  
BARAKA A, 1982, MIDDLE E J ANAESTHES, V6, P4
[4]  
CHAKRABORTY I, 2000, SENSOR ACTUAT A-PHYS, V83, P1
[5]  
COOMBS DW, 1991, ANESTH ANALG, V73, P226
[6]  
CRAWFORD JS, 1980, BRIT MED J, V281, P6248
[7]   Intrathecal drug delivery for treatment of chronic low back pain: Report from the National Outcomes Registry for Low Back Pain [J].
Deer, T ;
Chapple, I ;
Classen, A ;
Javery, K ;
Stoker, V ;
Tonder, L ;
Burchiel, K .
PAIN MEDICINE, 2004, 5 (01) :6-13
[8]  
DEER TR, 2002, SPINE J, V2, P4
[9]  
DUBOIS P, 2001, P SPIE INT SOC OPTIC, V4560
[10]   NORMALLY CLOSED MICROVALVE AND MICROPUMP FABRICATED ON A SILICON-WAFER [J].
ESASHI, M ;
SHOJI, S ;
NAKANO, A .
SENSORS AND ACTUATORS, 1989, 20 (1-2) :163-169