Routine use of a commercial test, GLABRATA RTT, for rapid identification of Candida glabrata in six laboratories
被引:13
作者:
Freydiere, AM
Perry, JD
Faure, O
Willinger, B
Tortorano, AM
Nicholson, A
Peman, J
Verweij, PE
机构:
[1] Hop Debrousse, Hosp Civils Lyon, Bacteriol Lab, F-69322 Lyon 05, France
[2] Ctr Hosp Univ, Grenoble, France
[3] Freeman Rd Hosp, Newcastle Upon Tyne NE7 7DN, Tyne & Wear, England
[4] Med Univ Vienna, Div Clin Microbiol, Inst Med Microbiol & Hyg, Vienna, Austria
[5] Univ Milan, Milan, Italy
[6] Univ Valencia, Hosp La Fe, Valencia, Spain
[7] Univ Nijmegen, Med Ctr, Dept Med Microbiol, Nijmegen, Netherlands
关键词:
D O I:
10.1128/JCM.42.10.4870-4872.2004
中图分类号:
Q93 [微生物学];
学科分类号:
071005 ;
100705 ;
摘要:
When evaluated in six clinical laboratories from six countries with 1,174 fresh isolates, including 715 Candida glabrata and 459 non-C. glabrata strains, GLABRATA RTT (Fumouze Diagnostics, Levallois Perret, France) yielded an overall sensitivity and an overall specificity of 95.8 and 98.9%, respectively. The results were consistent from one laboratory to another. The five false-positive results corresponded to C. parapsilosis (n = 2), C. tropicalis, C. guilliermondii, and C. lusitaniae. GLABRATA RTT allows a rapid, cost-effective, and reliable presumptive identification of C. glabrata.
引用
收藏
页码:4870 / 4872
页数:3