Dual Energy Subtraction Digital Radiography Improves Performance of a Next Generation Computer-aided Detection Program

被引:16
作者
Balkman, Jason D. [1 ]
Mehandru, Sonali [1 ]
DuPont, Elena [1 ]
Novak, Ronald D. [1 ]
Gilkeson, Robert C. [1 ]
机构
[1] Univ Hosp Cleveland, Dept Radiol, Cleveland, OH 44106 USA
关键词
dual energy subtraction; digital radiography; computer-aided detection; pulmonary nodules; FLAT-PANEL DETECTOR; CHEST RADIOGRAPHS; LUNG NODULES; DIAGNOSIS; CANCER; RADIOLOGISTS; SCHEME;
D O I
10.1097/RTI.0b013e3181aa34ed
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
Purpose: Computer-aided detection (CAD) has shown potential to assist physicians in the detection of lung nodules on chest radiographs, but widespread acceptance has been stymied by high false-positive rates. Few studies have examined the potential for dual energy subtraction (DES) to improve CAD performance. Materials and Methods: Institutional review board approval was obtained, the requirement for informed consent was waived because the study was retrospective, and practices conformed to Health Insurance Portability and Accountability Act regulations. The CAD program was applied retrospectively to dual energy posteroanterior (PA) chest radiographs of 36 patients (17 women, 19 men, mean age 69 y) with 48 pathology proven lung nodules. Results were analyzed to determine the stand-alone CAD program false-positive rates, and sensitivity by nodule subtlety and location. Statistical analysis was performed using the chi(2) or Fisher exact tests for independence of sensitivities between standard PA and DES radiography. Differences in the mean false-positives per image (FPPI) between radiographic modalities were determined using the paired Students t test, and bootstrap confidence intervals were obtained to confirm results. Results: The sensitivity of the CAD program with the standard PA was 46% (22 of 48 nodules) compared with 67% (32 of 48 nodules) using the DES soft tissue or bone-subtracted view (P = 0.064). The average number of FPPI identified by CAD was significantly lower using DES (FPPI(soft) (tissue) = 1.64) when compared with the standard PA chest radiograph (FPPI(PA) = 2.39) (P < 0.01). Conclusions: DES has the potential to improve stand-alone CAD performance by both increasing sensitivity for certain subtle lung cancer lesions and decreasing overall CAD false-positive rates.
引用
收藏
页码:41 / 47
页数:7
相关论文
共 20 条
[1]
AUSTIN JH, 1992, RADIOLOGY, V182, P8
[2]
Comparison of radiologist and CAD performance in the detection of CT-confirmed subtle pulmonary nodules on digital chest radiographs [J].
Bley, Thorsten Alexander ;
Baumann, Tobias ;
Saueressig, Ulrich ;
Pache, Gregor ;
Treier, Markus ;
Schaefer, Oliver ;
Neitzel, Ulrich ;
Langer, Mathias ;
Kotter, Elmar .
INVESTIGATIVE RADIOLOGY, 2008, 43 (06) :343-348
[3]
Doi K., 1990, US patent, Patent No. [4,907,156, 4907156]
[4]
Computer-aided diagnosis in medical imaging: Historical review, current status and future potential [J].
Doi, Kunio .
COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 2007, 31 (4-5) :198-211
[5]
Imaging characteristics of an amorphous silicon flat-panel detector for digital chest radiography [J].
Floyd, CE ;
Warp, RJ ;
Dobbins, JT ;
Chotas, HG ;
Baydush, AH ;
Vargas-Voracek, R ;
Ravin, CE .
RADIOLOGY, 2001, 218 (03) :683-688
[6]
Performance of a 41X41-cm2 amorphous silicon flat panel x-ray detector for radiographic imaging applications [J].
Granfors, PR ;
Aufrichtig, R .
MEDICAL PHYSICS, 2000, 27 (06) :1324-1331
[7]
Improved detection of lung nodules on chest radiographs using a commercial computer-aided diagnosis system [J].
Kakeda, S ;
Moriya, J ;
Sato, H ;
Aoki, T ;
Watanabe, H ;
Nakata, H ;
Oda, N ;
Katsuragawa, S ;
Yamamoto, K ;
Doi, K .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2004, 182 (02) :505-510
[8]
Effect of a computer-aided diagnosis scheme on radiologists' performance in detection of lung nodules on radiographs [J].
Kobayashi, T ;
Xu, XW ;
MacMahon, H ;
Metz, CE ;
Doi, K .
RADIOLOGY, 1996, 199 (03) :843-848
[10]
Improved detection of small lung cancers with dual-energy subtraction chest radiography [J].
Li, Feng ;
Engelmann, Roger ;
Doi, Kunio ;
MacMahon, Heber .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2008, 190 (04) :886-891