Recording of electrical brain activity in a magnetic resonance environment:: Distorting effects of the static magnetic field

被引:46
作者
Müri, RM
Felblinger, J
Rösler, KM
Jung, B
Hess, CW
Boesch, C
机构
[1] Univ Bern, Dept Neurol, CH-3010 Bern, Switzerland
[2] Univ Bern, Dept Magnet Resonance Spect & Methodol, CH-3010 Bern, Switzerland
[3] Univ Bern, Inselspital, CH-3010 Bern, Switzerland
关键词
electroencephalogram; visually evoked potentials; magnetic resonance; patient monitoring;
D O I
10.1002/mrm.1910390105
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The technical limitations of electroencephalography (EEG) and flashed visually-evoked potentials (VEP) recordings in the static magnetic field of the MR system were systematically studied. A main artifact occurring in the magnetic field was found to be correlated with the heart cycle and had amplitudes in the range of EEG and VEP signals. For VEP recordings, a substantial reduction of this effect was achieved by subtraction of the averaged artifact from the averaged composed signal (VEP and artifact) resulting in the VEP signal alone. However, for continuous EEG recordings, there is no such solution, since the observed effect is not sufficiently constant in amplitude, and the standard deviation of the amplitude of the effect is often larger than the EEG amplitude.
引用
收藏
页码:18 / 22
页数:5
相关论文
共 13 条
[11]   MAGNETOHYDRODYNAMICS OF BLOOD-FLOW [J].
KELTNER, JR ;
ROOS, MS ;
BRAKEMAN, PR ;
BUDINGER, TF .
MAGNETIC RESONANCE IN MEDICINE, 1990, 16 (01) :139-149
[12]  
Korchevskii E M, 1965, Biofizika, V10, P371
[13]  
VARDANYA.VA, 1973, BIOFIZIKA+, V18, P491