Comparison of the reproducibility in migration times between a constant-current and a constant-voltage mode of operation in capillary zone electrophoresis

被引:10
作者
Kurosu, Y [1 ]
Satou, Y [1 ]
Shisa, Y [1 ]
Iwata, T [1 ]
机构
[1] Jasco Tech Res Labs Corp, Hachioji, Tokyo 192, Japan
关键词
migration times; capillary electrophoresis; reproducibility;
D O I
10.1016/S0021-9673(97)01221-1
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A comparative study of the reproducibility in migration times between constant-current (CC) and a constant-voltage (CV) mode of operation in capillary zone electrophoresis was carried out. We found that the CC mode gave better reproducibility than the CV mode in both successive injections (5 or 6 times) and day-to-day analyses (16 measurements throughout 3 days). (C) 1998 Elsevier Science B.V.