In situ assembled mass-transport controlling micromembranes and their application in implanted amperometric glucose sensors

被引:72
作者
Chen, T
Friedman, KA
Lei, I
Heller, A [1 ]
机构
[1] Univ Texas, Dept Chem Engn, Austin, TX 78712 USA
[2] Univ Texas, Texas Mat Inst, Austin, TX 78712 USA
关键词
D O I
10.1021/ac000348c
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Micromembranes were assembled by sequentially chemisorbing polyanions and polycations on miniature (5 x 10(-4) cm(2)) enzyme electrodes. The sequential chemisorption process allowed the simultaneous tailoring of their sensitivity, dynamic range, drift, and selectivity. When assembled on tips of 250-mu m-diameter gold wires coated with redox polymer-"wired" glucose oxidase, they allowed tailoring of the glucose electrodes for >2 nA/mM sensitivity; 0-30 mM dynamic range; drift of less than or equal to 5% per 24 h at 37 degrees C at 15 mM glucose concentration; and 15% current increment by the combination of 0.1 mM ascorbate, 0.2 mM acetaminophen, and 0.5 mM urate, The membranes also retained transition metal ions that bound to and damaged the redox polymer "wiring" the enzyme. The electrodes were tested in the jugular veins and in the intrascapular subcutaneous region of anaesthetized and heparinized nondiabetic Sprague-Dawley rats, in which rapid changes of glycemia were forced by intravenous injections of glucose and insulin. After one-point in vivo calibration of the electrodes, all of the 152 data points were clinically accurate when it was assumed that after insulin injection the glycemia in the subcutaneous fluid lags by 9 min behind that of blood withdrawn from the insulin-injected vein.
引用
收藏
页码:3757 / 3763
页数:7
相关论文
共 29 条
  • [1] APPLICATION OF CHRONIC INTRAVASCULAR BLOOD-GLUCOSE SENSOR IN DOGS
    ARMOUR, JC
    LUCISANO, JY
    MCKEAN, BD
    GOUGH, DA
    [J]. DIABETES, 1990, 39 (12) : 1519 - 1526
  • [2] MICRODIALYSIS MEASUREMENT OF THE ABSOLUTE GLUCOSE-CONCENTRATION IN SUBCUTANEOUS ADIPOSE-TISSUE ALLOWING GLUCOSE MONITORING IN DIABETIC-PATIENTS
    BOLINDER, J
    UNGERSTEDT, U
    ARNER, P
    [J]. DIABETOLOGIA, 1992, 35 (12) : 1177 - 1180
  • [3] ELECTRODE SYSTEMS FOR CONTINUOUS MONITORING IN CARDIOVASCULAR SURGERY
    CLARK, LC
    LYONS, C
    [J]. ANNALS OF THE NEW YORK ACADEMY OF SCIENCES, 1962, 102 (01) : 29 - &
  • [4] EVALUATING CLINICAL ACCURACY OF SYSTEMS FOR SELF-MONITORING OF BLOOD-GLUCOSE
    CLARKE, WL
    COX, D
    GONDERFREDERICK, LA
    CARTER, W
    POHL, SL
    [J]. DIABETES CARE, 1987, 10 (05) : 622 - 628
  • [5] CSOREGI E, 1995, ANAL CHEM, V67, P1240
  • [6] CSOREGI E, 1994, ANAL CHEM, V66, P3131
  • [7] Fuzzy nanoassemblies: Toward layered polymeric multicomposites
    Decher, G
    [J]. SCIENCE, 1997, 277 (5330) : 1232 - 1237
  • [8] ELECTRICAL COMMUNICATION BETWEEN REDOX CENTERS OF GLUCOSE-OXIDASE AND ELECTRODES VIA ELECTROSTATICALLY AND COVALENTLY BOUND REDOX POLYMERS
    DEGANI, Y
    HELLER, A
    [J]. JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1989, 111 (06) : 2357 - 2358
  • [9] ASSESSMENT OF SUBCUTANEOUS GLUCOSE-CONCENTRATION - VALIDATION OF THE WICK TECHNIQUE AS A REFERENCE FOR IMPLANTED ELECTROCHEMICAL SENSORS IN NORMAL AND DIABETIC DOGS
    FISCHER, U
    ERTLE, R
    ABEL, P
    REBRIN, K
    BRUNSTEIN, E
    VONDORSCHE, HH
    FREYSE, EJ
    [J]. DIABETOLOGIA, 1987, 30 (12) : 940 - 945
  • [10] EVALUATION OF A SUBCUTANEOUS GLUCOSE SENSOR OUT TO 3 MONTHS IN A DOG-MODEL
    GILLIGAN, BJ
    SHULTS, MC
    RHODES, RK
    UPDIKE, SJ
    [J]. DIABETES CARE, 1994, 17 (08) : 882 - 887