Accuracy of the continuous glucose monitoring system in inpatient and outpatient conditions

被引:37
作者
Djakouré-Platonoff, C [1 ]
Radermercker, R [1 ]
Reach, G [1 ]
Slama, G [1 ]
Selam, JL [1 ]
机构
[1] Hop Hotel Dieu, Serv Diabetol, F-75181 Paris 4, France
关键词
glucose sensor; hypoglycemia; articificial pancreas;
D O I
10.1016/S1262-3636(07)70023-X
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: To evaluate in inpatient and outpatient conditions, using only non-calibration data, the accuracy of the Continuous Glucose Monitoring System,(CG-MS). Research design and methods: Twelve 21 and 20 type 1 diabetic patients participated in 3- day inhospital, 6-day ambulatory (i.e., 2 sensors worn consecutively) and 3-day ambulatory studies. Capillary glucose tests (7-8/day) served as comparisons. Pairs of data (excluding the 4 daily pairs used automatically by the CGMS for calibrations) were analyzed using correlation coefficient, error-grid analysis, and mean of absolute differences. Results: Two third and half of the sensing days were qualified as "optimal", i.e., interpretable by the CGMS program in the inpatient and outpatient studies, respectively. Correlation coefficient was 0.92, 0.81 and 0.73 in the inpatient, 6 day outpatient and 3-day outpatient studies. 98%, 93% and 96% of the data tell in the A or B error-grid "clinically acceptable" zones. Mean absolute differences between sensor and glucometer values were 25 +/- 2, 34 +/- 5 and 32 +/- 5 mg/dl. Conclusion: CGMS may be used for assessment of glycemic trends, though accuracy is more in the range of glucometers; than of laboratory tests. Inhospital utilisation improved accuracy and 6-day rather than 3-day sessions may be more appropriate to evaluate day to day reproducibility of glucose profile.
引用
收藏
页码:159 / 162
页数:4
相关论文
共 14 条
[1]  
[Anonymous], **NON-TRADITIONAL**, DOI DOI 10.1089/15209150050214087
[2]   STATISTICS NOTES .1. CORRELATION, REGRESSION, AND REPEATED DATA [J].
BLAND, JM ;
ALTMAN, DG .
BRITISH MEDICAL JOURNAL, 1994, 308 (6933) :896-896
[3]   Continuous glucose monitoring used to adjust diabetes therapy improves glycosylated hemoglobin: a pilot study [J].
Bode, BW ;
Gross, TM ;
Thornton, KR ;
Mastrototaro, JJ .
DIABETES RESEARCH AND CLINICAL PRACTICE, 1999, 46 (03) :183-190
[4]   Prevention of hypoglycemia using risk assessment with a continuous glucose monitoring system [J].
Choleau, C ;
Dokladal, P ;
Klein, JC ;
Ward, WK ;
Wilson, GS ;
Reach, G .
DIABETES, 2002, 51 (11) :3263-3273
[5]   EVALUATING CLINICAL ACCURACY OF SYSTEMS FOR SELF-MONITORING OF BLOOD-GLUCOSE [J].
CLARKE, WL ;
COX, D ;
GONDERFREDERICK, LA ;
CARTER, W ;
POHL, SL .
DIABETES CARE, 1987, 10 (05) :622-628
[6]   Correlation at fingerstick blood glucose measurements with GlucoWatch biographer glucose results in young subjects with type 1 diabetes [J].
Garg, SK ;
Potts, RO ;
Ackerman, NR ;
Fermi, SJ ;
Tamada, JA ;
Chase, HP .
DIABETES CARE, 1999, 22 (10) :1708-1714
[7]  
Gross T M, 2000, Diabetes Technol Ther, V2, P49, DOI 10.1089/152091500316737
[8]  
GROSS TM, 2000, DIABETES TECHNOL THE, V2, P27
[9]  
HIRSCH IB, 2000, DIABETES TECHNOL THE, V2, P581
[10]  
Mastrototaro J J, 2000, Diabetes Technol Ther, V2 Suppl 1, pS13, DOI 10.1089/15209150050214078