Quality assurance by systematic in vivo dosimetry: results on a large cohort of patients

被引:48
作者
Fiorino, C
Corletto, D
Mangili, P
Broggi, S
Bonini, A
Cattaneo, GM
Parisi, R
Rosso, A
Signorotto, P
Villa, E
Calandrino, R
机构
[1] Hosp San Raffaele, Serv Fis Sanit, I-20132 Milan, Italy
[2] HS Raffaele, Serv Radiochemioterapia, Milan, Italy
关键词
in vivo dosimetry; quality assurance; errors in radiotherapy;
D O I
10.1016/S0167-8140(00)00195-X
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: In vivo dosimetry is widely considered to be an important tool for quality assurance in external radiotherapy. Introduction: In this study we report on our experience over more than 4 years in systematic in vivo dosimetry with diodes. Materials and methods: From November '94 an in vivo entrance dosimetry check was performed for every new patient irradiated at one of our treatment units (Linac 6/100, 6 MV X-rays). Diodes were calibrated in terms of entrance dose; appropriate correction factors had been previously assessed (taking SSDs, field width, wedge, oblique incidence and blocking tray into account) and were individually applied to in vivo diode readings. The in vivo measured entrance dose was compared with the expected one, with a 5% action level; if a larger deviation was found, all treatment parameters were verified, and the in vivo dosimetry check was repeated. During the period November '94-May '99, 2824 measurements on 1433 patients were collected. Results: Nine out of 1433 (0.63%) serious systematic errors (leading to a 5% or more on the delivered dose to the PTV) were detected by in vivo dosimetry; four out of nine would produce a 10% or more error if not detected. The rate of serious systematic errors detected by an independent check of treatment chart and MU calculation was found to be 1.5%, showing that less than 1/3 of the errors escapes this check. One hundred and twelve out of 1433 (7.8%) patients had more than one check: the rate of second checks was significantly higher for breast patients (31/250, 12.4%) against non-breast patients (81/1183, 6.8%, P = 0.003). A number of patients demonstrated a persistent relatively large error even after two or more checks. For almost all patients the cause of the deviation was assessed; the most frequent cause was the difficulty in correctly positioning the patient and/or the diode. When analyzing the distribution of the deviations between measured and expected entrance doses (excluding first checks in the case of repetition of the in vivo dosimetry control) the mean deviation was 0.4% with a standard deviation equal to 3.0%. The rates of deviations larger than 5 and 7% were 9.9 and 2.6%, respectively. When considering the same data taking the average deviation in the case of opposed beams, the SD became 2.6% and the rates of deviations larger than 5 and 7%, respectively, 5.2 and 0.8%. When dividing the beams according to their orientation, significantly higher rates of large deviations (>5 and 7%) were found for oblique and posterior-anterior (PA) fields against lateral and anterior-posterior (AP) fields (P < 0.05). Similarly, higher rates of large deviations were found for wedged fields against unwedged fields (P < 0.03) and for blocked fields against unblocked fields (P < 0.01). When dividing the data according to the anatomical district, accuracy was worse for breast (mean deviation 0.1%, 1 SD: 3.5%) and neck AP-PA fields (mean deviation 1%, 1 SD: 3,4%). Better accuracy was found for vertebrae (0.1%, 1 SD 2.1%) and brain patients (-0.7%, 1 SD: 2.6%). During the considered period, in vivo dosimetry was also able to promptly detect a systematic error caused by a wrong resetting of the simulator height couch indicator, with a consequent error in the estimate of patient thickness of about 4 cm. Conclusions: In our experience, systematic in vivo dosimetry demonstrated to be a valid tool for quality assurance, both in detecting systematic errors which may escape the data transfer/MU calculation check and in giving an effective way of estimating the accuracy of treatment delivery. (C) 2000 Elsevier Science Ireland Ltd. All rights reserved.
引用
收藏
页码:85 / 95
页数:11
相关论文
共 35 条
[1]   QUALITY ASSURANCE IN RADIATION-THERAPY - SYSTEMATIC EVALUATION OF ERRORS DURING THE TREATMENT EXECUTION [J].
BLANCO, S ;
LOPEZBOTE, MA ;
DESCO, M .
RADIOTHERAPY AND ONCOLOGY, 1987, 8 (03) :253-261
[2]   Detection of systematic errors in external radiotherapy before treatment delivery [J].
Calandrino, R ;
Cattaneo, GM ;
Fiorino, C ;
Longobardi, B ;
Mangili, P ;
Signorotto, P .
RADIOTHERAPY AND ONCOLOGY, 1997, 45 (03) :271-274
[3]   HUMAN ERRORS IN THE CALCULATION OF MONITOR UNITS IN CLINICAL RADIOTHERAPY PRACTICE [J].
CALANDRINO, R ;
CATTANEO, GM ;
DELVECCHIO, A ;
FIORINO, C ;
LONGOBARDI, B ;
SIGNOROTTO, P .
RADIOTHERAPY AND ONCOLOGY, 1993, 28 (01) :86-88
[4]   QUALITY-CONTROL IN THE CONSERVATIVE TREATMENT OF BREAST-CANCER - PATIENT DOSIMETRY USING SILICON DETECTORS [J].
CIOCCA, M ;
LANDONI, L ;
ITALIA, C ;
MONTANARO, P ;
CANESI, P ;
VALDAGNI, R .
RADIOTHERAPY AND ONCOLOGY, 1991, 22 (04) :304-307
[5]   Quality assurance in radiation oncology. A study of feasibility and impact on action levels of an in vivo dosimetry program during breast cancer irradiation [J].
Cozzi, L ;
Fogliata-Cozzi, A .
RADIOTHERAPY AND ONCOLOGY, 1998, 47 (01) :29-36
[6]   Accuracy of two- and three-dimensional photon dose calculation for tangential irradiation of the breast [J].
Davis, JB ;
Pfafflin, A ;
Cozzi, AF .
RADIOTHERAPY AND ONCOLOGY, 1997, 42 (03) :245-248
[7]   An independent check of treatment plan, prescription and dose calculation as a QA procedure [J].
Duggan, L ;
Kron, T ;
Howlett, S ;
Skov, A ;
OBrien, P .
RADIOTHERAPY AND ONCOLOGY, 1997, 42 (03) :297-301
[8]   In vivo dosimetry during external photon beam radiotherapy [J].
Essers, M ;
Mijnheer, BJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1999, 43 (02) :245-259
[9]   The accuracy of CT-based inhomogeneity corrections and in vivo dosimetry for the treatment of lung cancer [J].
Essers, M ;
Lanson, JH ;
Leunens, G ;
Schnabel, T ;
Mijnheer, BJ .
RADIOTHERAPY AND ONCOLOGY, 1995, 37 (03) :199-208
[10]   IN-VIVO DOSIMETRY DURING CONFORMAL THERAPY OF PROSTATIC-CANCER [J].
ESSERS, M ;
LANSON, JH ;
MIJNHEER, BJ .
RADIOTHERAPY AND ONCOLOGY, 1993, 29 (02) :271-279