Does the endovascular repair of aortoiliac aneurysms pose a radiation safety hazard to vascular surgeons?

被引:55
作者
Lipsitz, EC
Veith, FJ
Ohki, T
Heller, S
Wain, RA
Suggs, WD
Lee, JC
Kwei, S
Goldstein, K
Rabin, T
Chang, D
Mehta, M
机构
[1] Montefiore Med Ctr, Div Vasc Surg, Vasc Lab, Dept Surg, Bronx, NY 10467 USA
[2] Albert Einstein Coll Med, Bronx, NY 10467 USA
关键词
D O I
10.1067/mva.2000.110053
中图分类号
R61 [外科手术学];
学科分类号
摘要
Objectives: Endovascular aortoiliac aneurysm (EAIA) repair uses substantial fluoroscopic guidance that requires considerable radiation exposure. Doses were determined for a team of three vascular surgeons performing 47 consecutive EAIA repairs over a 1-year period to determine whether this exposure constitutes a radiation hazard. Methods: Twenty-nine surgeon-made aortounifemoral devices and Is bifurcated devices were used. Three surgeons wore dosimeters (1) on the waist, under a lead apron; (2) on the waist, outside a lead apron; (3) on the collar; and (4) on the left ring finger. Dosimeters were also placed around the operating table and room to evaluate the patient, other personnel, and ambient doses. Exposures were compared With standards of the International Commission on Radiological Protection (ICRP). Results: Total fluoroscopy time was 30.9 hours (1852 minutes; mean, 39.4 minutes per case). Yearly total effective body doses for all surgeons (under lead) were below the 20 mSv/y occupational exposure limit of the ICRP. Outside lead doses for two surgeons approximated recommended limits. Lead aprons attenuated 85% to 91% of the dose. Ring doses and calculated eye doses were within the ICRT! exposure limits. Patient skin doses averaged 360 mSv per case (range, 120-860 mSv). The ambient (> 3 m from the source) operating room dose was 1.06 mSv/y. Conclusions: Although the total effective body doses under lead fell within established ICRP occupational exposure limits, they are not negligible. Because radiation exposure is cumulative and endovascular procedures are becoming more common, individuals performing these procedures must carefully monitor their exposure. Our results indicate that a team of surgeons can perform 386 hours of fluoroscopy per year or 587 EAIA repairs per year and remain within occupational exposure limits. Individuals who perform these procedures should actively monitor their effective doses and educate personnel in methods for reducing exposure.
引用
收藏
页码:704 / 709
页数:6
相关论文
共 19 条
[1]  
Aldridge HE, 1997, CAN J CARDIOL, V13, P459
[2]  
[Anonymous], 1996, Ann ICRP, V26, P1
[3]   RADIATION EXPOSURE TO ANGIOGRAPHERS UNDER DIFFERENT FLUOROSCOPIC IMAGING CONDITIONS [J].
BOONE, JM ;
LEVIN, DC .
RADIOLOGY, 1991, 180 (03) :861-865
[4]   Measurement of radiation dose received by the hands and thyroid of staff performing gridless fluoroscopic procedures in children [J].
Coakley, KS ;
Ratcliffe, J ;
Masel, J .
BRITISH JOURNAL OF RADIOLOGY, 1997, 70 (837) :933-936
[5]   THE CASE FOR RADIOPROTECTIVE EYEWEAR FACEWEAR - PRACTICAL IMPLICATIONS AND SUGGESTIONS [J].
COUSIN, AJ ;
LAWDAHL, RB ;
CHAKRABORTY, DP ;
KOEHLER, RE .
INVESTIGATIVE RADIOLOGY, 1987, 22 (08) :688-692
[6]   Primary beam exposure outside the fluoroscopic field of view [J].
Granger, WE ;
Bednarek, DR ;
Rudin, S .
MEDICAL PHYSICS, 1997, 24 (05) :703-707
[7]   Radiation exposure to patients and personnel during interventional ERCP at a teaching institution [J].
Heyd, RL ;
Kopecky, KK ;
Sherman, S ;
Lehman, GA ;
Stockberger, SM .
GASTROINTESTINAL ENDOSCOPY, 1996, 44 (03) :287-292
[8]   AUDIBLE RADIATION MONITORS - THE VALUE IN REDUCING RADIATION EXPOSURE TO FLUOROSCOPY PERSONNEL [J].
HOUGH, DM ;
BRADY, A ;
STEVENSON, GW .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1993, 160 (02) :407-408
[9]   EFFECTIVENESS OF LEAD APRONS IN VASCULAR RADIOLOGY - RESULTS OF CLINICAL MEASUREMENTS [J].
KICKEN, PJ ;
BOS, AJJ .
RADIOLOGY, 1995, 197 (02) :473-478
[10]  
KUWAYAMA N, 1994, AM J NEURORADIOL, V15, P1801