Automatic Column Coupling System To Operate Chromatographic Supports Closer To Their Kinetic Performance Limit and To Enhance Method Development

被引:14
作者
Cabooter, Deirdre [1 ]
Decrop, Wim [2 ]
Eeltink, Sebastiaan [1 ,2 ]
Swart, Remco [2 ]
Ursem, Mario [2 ]
Lestremau, Francois [3 ]
Desmet, Gert [1 ]
机构
[1] Vrije Univ Brussel, Dept Chem Engn, B-1050 Brussels, Belgium
[2] Dionex Chem Corp, NL-1046 AA Amsterdam, Netherlands
[3] Pfizer Global Res & Dev, Sandwich CT13 9NJ, Kent, England
关键词
EFFICIENCY LIQUID-CHROMATOGRAPHY; METHOD DEVELOPMENT STRATEGY; CONVENTIONAL COLUMNS; PLOT METHOD; PHASE; SEPARATION; RETENTION; INSTRUMENTATION; REPRESENTATION; TEMPERATURE;
D O I
10.1021/ac902404v
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new hardware solution is proposed that allows one to automatically change the length of a chromatographic bed. The setup is based on the serial coupling of chromatographic columns using two rotor-stator valves (with N positions, N + 1 ports). Despite the use of a prototype setup requiring rather long connection tubing, only 9% loss in efficiency is observed for compounds with a retention factor above 4 compared to the efficiency expected on the basis of the individual column results. It has been demonstrated for a number of isocratic and gradient separations that the system allows one to realize considerable analysis time savings by adapting the total column length to the specific sample requirements and/or to the stage of method development wherein one is working. During method development, a separation on a short column length can first be used to rapidly gain insight into the composition of the sample, leaving fewer runs to be done on a column of maximal length (offering efficiencies that are inaccessible with individual column systems). The ease with which information can be obtained on columns of different lengths can furthermore be exploited for screening purposes to detect coeluting components in a stage wherein they still appear completely unresolved (i.e., have a resolution well below R-s = 0.5).
引用
收藏
页码:1054 / 1065
页数:12
相关论文
共 30 条
[21]   High-performance liquid chromatography of basic compounds - Problems, possible solutions and tests of reversed-phase columns [J].
McCalley, DV ;
Brereton, RG .
JOURNAL OF CHROMATOGRAPHY A, 1998, 828 (1-2) :407-420
[22]   Computerized design of separation strategies by reversed-phase liquid chromatography: development of DryLab software [J].
Molnar, I .
JOURNAL OF CHROMATOGRAPHY A, 2002, 965 (1-2) :175-194
[23]  
Neue U., 1997, HPLC Columns. Theory, P43
[24]  
NEUE UD, 2003, HDB ANAL SEPARATIONS, P185
[25]   Performance of idealized column structures under high pressure [J].
Neue, Uwe D. ;
Kele, Marianna .
JOURNAL OF CHROMATOGRAPHY A, 2007, 1149 (02) :236-244
[27]   MODE OF OPERATION AND PERFORMANCE-CHARACTERISTICS OF MICROBORE COLUMNS FOR USE IN LIQUID-CHROMATOGRAPHY [J].
SCOTT, RPW ;
KUCERA, P .
JOURNAL OF CHROMATOGRAPHY, 1979, 169 (FEB) :51-72
[28]  
Snyder L.R., 1997, PRACTICAL HPLC METHO, V2nd
[29]   Retention of ionisable compounds on high-performance liquid chromatography XVI.: Estimation of retention with acetonitrile/water mobile phases from aqueous buffer pH and analyte pKa [J].
Subirats, Xavier ;
Bosch, Elisabeth ;
Roses, Marti .
JOURNAL OF CHROMATOGRAPHY A, 2006, 1121 (02) :170-177
[30]   Efficient method development strategy for challenging separation of pharmaceutical molecules using advanced chromatographic technologies [J].
Xiao, Kang Ping ;
Xiong, Yuan ;
Liu, Fang Zhu ;
Rustum, Abu M. .
JOURNAL OF CHROMATOGRAPHY A, 2007, 1163 (1-2) :145-156