Numerical simulation of SAR and B1-field inhomogeneity of shielded RF coils loaded with the human head

被引:108
作者
Chen, J
Feng, ZM
Jin, JM
机构
[1] Univ Illinois, Dept Elect & Comp Engn, Electromagnet Lab, Urbana, IL 61801 USA
[2] Univ Illinois, Dept Elect & Comp Engn, Magnet Resonance Engn Lab, Urbana, IL 61801 USA
关键词
magnetic resonance imaging; numerical method; radio-frequency coil; radio-frequency field; specific absorption rate;
D O I
10.1109/10.668756
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The finite-difference time-domain (FDTD) method is combined with the method of moments (MoM) to compute the electromagnetic fields of shielded radio frequency (RF) coils loaded with an anatomically accurate model of a human head for high-frequency magnetic resonance imaging (MRI) applications. The combined method can predict both the specific energy absorption rate (SAR) and the magnetic field (known as the B-1 field) excited by any RF coils. Results for SAR and B-1 field distribution, excited by shielded and end-capped birdcage coils, are calculated at 64, 128, 171, and 256 MHz, The results show that the value of SAR increases when the frequency of the B-1 field increases and the B-1 field exhibits a strong inhomogeneity at high frequencies.
引用
收藏
页码:650 / 659
页数:10
相关论文
共 31 条
[11]   AN EFFICIENT, HIGHLY HOMOGENEOUS RADIOFREQUENCY COIL FOR WHOLE-BODY NMR IMAGING AT 1.5-T [J].
HAYES, CE ;
EDELSTEIN, WA ;
SCHENCK, JF ;
MUELLER, OM ;
EASH, M .
JOURNAL OF MAGNETIC RESONANCE, 1985, 63 (03) :622-628
[12]  
HAYES CE, 1986, SMRM WORKS PROGR, P39
[13]   ON THE FIELD INHOMOGENEITY OF A BIRDCAGE COIL [J].
JIN, JM ;
SHEN, G ;
PERKINS, T .
MAGNETIC RESONANCE IN MEDICINE, 1994, 32 (03) :418-422
[14]   On the SAR and field inhomogeneity of birdcage coils loaded with the human head [J].
Jin, JM ;
Chen, J .
MAGNETIC RESONANCE IN MEDICINE, 1997, 38 (06) :953-963
[15]   Computation of electromagnetic fields for high-frequency magnetic resonance imaging applications [J].
Jin, JM ;
Chen, J ;
Chew, WC ;
Gan, H ;
Magin, RL ;
Dimbylow, PJ .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (12) :2719-2738
[16]   A TECHNIQUE FOR DOUBLE RESONANT OPERATION OF BIRDCAGE IMAGING COILS [J].
JOSEPH, PM ;
LU, DF .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1989, 8 (03) :286-294
[17]   ELECTROMAGNETIC-FIELDS OF SURFACE COIL INVIVO NMR AT HIGH-FREQUENCIES [J].
KELTNER, JR ;
CARLSON, JW ;
ROOS, MS ;
WONG, STS ;
WONG, TL ;
BUDINGER, TF .
MAGNETIC RESONANCE IN MEDICINE, 1991, 22 (02) :467-480
[18]  
KUNZ KS, 1994, FINITE DIFFERENCE TI
[19]  
MANSFIELD P, 1982, ADV MAGNETIC RESON S, V2
[20]  
OCHI H, 1992, P SMRM 11 ANN M BERL, P4021