Diffusion-weighted MR imaging of the abdomen with pulse triggering.

被引:6
作者
Mürtz, P
Pauleit, D
Träber, F
Gieseke, J
Kreft, BP
van den Brink, JS
Schild, HH
机构
[1] Univ Bonn, Radiol Klin, D-53105 Bonn, Germany
[2] Philips Med Syst, Best, Netherlands
来源
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN | 2000年 / 172卷 / 07期
关键词
magnetic resonance imaging; abdomen; diffusion weighted images; apparent diffusion coefficients; motion artifacts;
D O I
10.1055/s-2000-4667
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The aim of this work was to reduce the influence of motion on diffusion-weighted MR images of the abdomen by pulse triggering of single-shot sequences. Methods: Five healthy volunteers were examined both without and with finger pulse-triggering of a diffusion-weighted single-shot echo planar MR imaging sequence at 1.5 T. Series of diffusion-weighted images were acquired at different phases of the cardiac cycle by varying the time delay between finger pulse and sequence acquisition. The measurements were repeated three times. The diffusion weighted images were analysed by measuring the signal intensities and by determining the ADC values within the spleen, kidney and liver. Results: The magnitude of motion artifacts on diffusion weighted images shows a strong dependence on the trigger delay. The optimum trigger delay is found to be between 500 and 600 ms. For these values the abdominal organs appear homogeneous on all diffusion weighted images and the strongest signal intensities are defected. At optimum triggering the accuracy of the apparent diffusion coefficients is up to 10 times better than without triggering. Moreover, the standard deviation of the repeated measurements is smaller than 12% for all volunteers and for all organs. Without triggering the standard deviation is larger by a factor of 4 on average. Conclusion: Pulse triggering of singleshot sequences leads to significant reduction of motion related artifacts on diffusion weighted images of the abdomen and provides more accurate and reproducible ADC values.
引用
收藏
页码:587 / 590
页数:4
相关论文
共 7 条
[1]  
Gray L, 1998, Magn Reson Imaging Clin N Am, V6, P125
[2]   Diffusion-weighted MR imaging with single-shot echo-planar imaging in the upper abdomen: preliminary clinical experience in 61 patients [J].
Ichikawa, T ;
Haradome, H ;
Hachiya, J ;
Nitatori, T ;
Araki, T .
ABDOMINAL IMAGING, 1999, 24 (05) :456-461
[3]   Diffusion-weighted single-shot echoplanar MR imaging for liver disease [J].
Kim, T ;
Murakami, T ;
Takahashi, S ;
Hori, M ;
Tsuda, K ;
Nakamura, H .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1999, 173 (02) :393-398
[4]   IN-VIVO DIFFUSION MEASUREMENTS OF LIVER, KIDNEY, SPLEEN AND SACROSPINALIS MUSCLE USING AN ECHO-PLANAR MR-IMAGING SYSTEM IN HEALTHY-PERSONS [J].
MULLER, MF ;
PRASAD, PV ;
SIEWERT, B ;
EDELMAN, RR .
FORTSCHRITTE AUF DEM GEBIETE DER RONTGENSTRAHLEN UND DER NEUEN BILDGEBENDEN VERFAHREN, 1994, 161 (03) :233-236
[5]   Focal liver masses: Characterization with diffusion-weighted echo-planar MR imaging [J].
Namimoto, T ;
Yamashita, Y ;
Sumi, S ;
Tang, Y ;
Takahashi, M .
RADIOLOGY, 1997, 204 (03) :739-744
[6]   Diffusion coefficients in abdominal organs and hepatic lesions: Evaluation with intravoxel incoherent motion echo-planar MR imaging [J].
Yamada, I ;
Aung, W ;
Himeno, Y ;
Nakagawa, T ;
Shibuya, H .
RADIOLOGY, 1999, 210 (03) :617-623
[7]   Ultrafast MR imaging of the abdomen: Echo planar imaging and diffusion-weighted imaging [J].
Yamashita, Y ;
Tang, Y ;
Takahashi, M .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1998, 8 (02) :367-374