A new generation of magnetoencephalography: Room temperature measurements using optically-pumped magnetometers

被引:302
作者
Boto, Elena [1 ]
Meyer, Sofie S. [2 ]
Shah, Vishal [3 ]
Alem, Orang [3 ]
Knappe, Svenja [3 ]
Kruger, Peter [4 ]
Fromhold, T. Mark [4 ]
Lim, Mark [5 ]
Glover, Paul M. [1 ]
Morris, Peter G. [1 ]
Bowtell, Richard [1 ]
Barnes, Gareth R. [2 ]
Brookes, Matthew J. [1 ]
机构
[1] Univ Nottingham, Sch Phys & Astron, Sir Peter Mansfield Imaging Ctr, Univ Pk, Nottingham NG7 2RD, England
[2] UCL, Inst Neurol, Wellcome Trust, Ctr Neuroimaging, 12 Queen Sq, London WC1N 3BG, England
[3] QuSpin Inc, 2011 Cherry St,Unit 112, Louisville, CO 80027 USA
[4] Univ Nottingham, Sch Phys & Astron, Midlands Ultracold Atom Res Ctr, Univ Pk, Nottingham NG7 2RD, England
[5] Chalk Studios Ltd, 14 Windsor St, London N1 8QG, England
基金
英国医学研究理事会; 英国生物技术与生命科学研究理事会; 英国惠康基金; 英国工程与自然科学研究理事会;
关键词
ATOMIC MAGNETOMETER; MEG; SENSITIVITY;
D O I
10.1016/j.neuroimage.2017.01.034
中图分类号
Q189 [神经科学];
学科分类号
071006 ;
摘要
Advances in the field of quantum sensing mean that magnetic field sensors, operating at room temperature, are now able to achieve sensitivity similar to that of cryogenically cooled devices (SQUIDs). This means that room temperature magnetoencephalography (MEG), with a greatly increased flexibility of sensor placement can now be considered. Further, these new sensors can be placed directly on the scalp surface giving, theoretically, a large increase in the magnitude of the measured signal. Here, we present recordings made using a single optically-pumped magnetometer (OPM) in combination with a 3D-printed head-cast designed to accurately locate and orient the sensor relative to brain anatomy. Since our OPM is configured as a magnetometer it is highly sensitive to environmental interference. However, we show that this problem can be ameliorated via the use of simultaneous reference sensor recordings. Using median nerve stimulation, we show that the OPM can detect both evoked (phase-locked) and induced (non-phase-locked oscillatory) changes when placed over sensory cortex, with signals 4 times larger than equivalent SQUID measurements. Using source modelling, we show that our system allows localisation of the evoked response to somatosensory cortex. Further, source-space modelling shows that, with 13 sequential OPM measurements, source-space signal-to-noise ratio (SNR) is comparable to that from a 271-channel SQUID system. Our results highlight the opportunity presented by OPMs to generate uncooled, potentially low-cost, high SNR MEG systems.
引用
收藏
页码:404 / 414
页数:11
相关论文
共 44 条
  • [1] Magnetoencephalography of Epilepsy with a Microfabricated Atomic Magnetrode
    Alem, Orang
    Benison, Alex M.
    Barth, Daniel S.
    Kitching, John
    Knappe, Svenja
    [J]. JOURNAL OF NEUROSCIENCE, 2014, 34 (43) : 14324 - 14327
  • [2] On the Potential of a New Generation of Magnetometers for MEG: A Beamformer Simulation Study
    Boto, Elena
    Bowtell, Richard
    Kruger, Peter
    Fromhold, T. Mark
    Morris, Peter G.
    Meyer, Sofie S.
    Barnes, Gareth R.
    Brookes, Matthew J.
    [J]. PLOS ONE, 2016, 11 (08):
  • [3] Optimising experimental design for MEG beamformer imaging
    Brookes, Matthew J.
    Vrba, Jiri
    Robinson, Stephen E.
    Stevenson, Claire M.
    Peters, Andrew M.
    Barnes, Gareth R.
    Hillebrand, Arjan
    Morris, Peter G.
    [J]. NEUROIMAGE, 2008, 39 (04) : 1788 - 1802
  • [4] Investigating spatial specificity and data averaging in MEG
    Brookes, Matthew J.
    Zumer, Johanna M.
    Stevenson, Claire M.
    Hale, Joanne R.
    Barnes, Gareth R.
    Vrba, Jiri
    Morris, Peter G.
    [J]. NEUROIMAGE, 2010, 49 (01) : 525 - 538
  • [5] MEG studies of sensorimotor rhythms: A review
    Cheyne, Douglas Owen
    [J]. EXPERIMENTAL NEUROLOGY, 2013, 245 : 27 - 39
  • [6] Habituation: Events in the history of its characterization and linkage to synaptic depression. A new proposed kinetic criterion for its identification
    Christoffersen, GRJ
    [J]. PROGRESS IN NEUROBIOLOGY, 1997, 53 (01) : 45 - 66
  • [7] Ciesielski K. R., 2014, MAGNETOENCEPHALOGRAP
  • [9] Four-channel optically pumped atomic magnetometer for magnetoencephalography
    Colombo, Anthony P.
    Carter, Tony R.
    Borna, Amir
    Jau, Yuan-Yu
    Johnson, Cort N.
    Dagel, Amber L.
    Schwindt, Peter D. D.
    [J]. OPTICS EXPRESS, 2016, 24 (14): : 15403 - 15416
  • [10] Consequences of EEG electrode position error on ultimate beamformer source reconstruction performance
    Dalal, Sarang S.
    Rampp, Stefan
    Willomitzer, Florian
    Ettl, Svenja