Ultraviolet radiation dosimetry with radiochromic film

被引:33
作者
Butson, MJ
Cheung, T
Yu, PKN
Abbati, D
Greenoak, GE
机构
[1] Illawarra Canc Care Ctr, Dept Med Phys, Wollongong, NSW 2500, Australia
[2] Wollongong Hosp, Dept Nucl Med, Wollongong, NSW 2500, Australia
[3] City Univ Hong Kong, Dept Phys & Mat Sci, Hong Kong, Hong Kong, Peoples R China
[4] Univ Sydney, Dept Anim Sci, Sydney, NSW 2006, Australia
关键词
D O I
10.1088/0031-9155/45/7/311
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Radiochromic film is tested for its broad-band response to ultraviolet (UV) B (290-320 nm) and A (320 nm-400 nm), visible and infrared radiation produced by a solar simulator and examined for dosimetry in ultraviolet radiation. Results show that MD-55-2 radiochromic film in solar and fluorescent light sources responds almost exclusively to broad-band UVA radiation with negligible colouration from UVB, visible and low level infrared radiation. A second order polynomial function approximates the change in optical density at 660 nm wavelength for film colouration with exposure to UVA from white light fluorescent and solar UV with exposures measured with a dedicated UVA dosimeter. Using a double exposure technique as used in radiation dosimetry where the film is firstly irradiated to a known UV dose, radiochromic film can be used as a quantitative measure of UVA exposure.
引用
收藏
页码:1863 / 1868
页数:6
相关论文
共 17 条
[1]   An estimate of the total UV-B exposure for outdoor workers during a south-east Queensland summer [J].
Airey, DK ;
Wong, JCF ;
Fleming, RA ;
Meldrum, LR .
HEALTH PHYSICS, 1997, 72 (04) :544-549
[2]   SUNBURNING ULTRAVIOLET METER - DESIGN AND PERFORMANCE [J].
BERGER, DS .
PHOTOCHEMISTRY AND PHOTOBIOLOGY, 1976, 24 (06) :587-593
[3]   Measurement of off-axis and peripheral skin dose using radiochromic film [J].
Butson, MJ ;
Yu, PKN ;
Metcalfe, PE .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (09) :2647-2650
[4]   POSSIBLE DOSIMETER FOR ULTRAVIOLET-RADIATION [J].
DAVIS, A ;
DEANE, GHW ;
DIFFEY, BL .
NATURE, 1976, 261 (5556) :169-170
[5]  
DIFFEY BL, 1989, RAD MEASUREMENT PHOT
[6]   EFFECTS OF ULTRAVIOLET-IRRADIATION ON HUMAN SKIN-DERIVED EPIDERMAL-CELLS IN-VITRO [J].
DISSANAYAKE, NS ;
GREENOAK, GE ;
MASON, RS .
JOURNAL OF CELLULAR PHYSIOLOGY, 1993, 157 (01) :119-127
[7]  
HORNECK G, 1993, BIOL EFFECTS LIGHT 1, P501
[8]   Quantification of personal solar UV exposure of outdoor workers, indoor workers and adolescents at two locations in Southeast Queensland [J].
Kimlin, MG ;
Parisi, AV ;
Wong, JCF .
PHOTODERMATOLOGY PHOTOIMMUNOLOGY & PHOTOMEDICINE, 1998, 14 (01) :7-11
[9]   SENSITOMETRY OF THE RESPONSE OF A NEW RADIOCHROMIC FILM DOSIMETER TO GAMMA-RADIATION AND ELECTRON-BEAMS [J].
MCLAUGHLIN, WL ;
YUNDONG, C ;
SOARES, CG ;
MILLER, A ;
VANDYK, G ;
LEWIS, DF .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1991, 302 (01) :165-176
[10]   Dosimetric characteristics of an improved radiochromic film [J].
Meigooni, AS ;
Sanders, MF ;
Ibbott, GS ;
Szeglin, SR .
MEDICAL PHYSICS, 1996, 23 (11) :1883-1888