A disposable biosensor for organophosphorus nerve agents based on carbon nanotubes modified thick film strip electrode

被引:183
作者
Joshi, KA
Tang, J
Haddon, R
Wang, J
Chen, W
Mulchandani, A [1 ]
机构
[1] Univ Calif Riverside, Dept Chem & Environm Engn, Riverside, CA 92521 USA
[2] Univ Calif Riverside, Dept Chem, Riverside, CA 92521 USA
[3] New Mexico State Univ, Dept Chem & Biochem, Las Cruces, NM 88003 USA
关键词
carbon nanotubes; thick film strip electrodes; biosensor; organophosphorus insecticides; acetylcholinesterase; paraoxons;
D O I
10.1002/elan.200403118
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A disposable biosensor based on acetylcholinesterase-functionalized acid purified multi-wall carbon nanotubes (CNTs) modified thick film strip electrode for organophosphorus (OP) insecticides was developed. The degree of inhibition of the enzyme acteylcholinesterase (AChE) by OP compounds was determined by measuring the electrooxidation current of the thiocholine generated by the AChE catalyzed hydrolysis of acteylthiocholine (ATCh). The large surface area and electro-catalytic activity of carbon nanotubes lowered the overpotential for thiocholine oxidation to 200 mV (vs. Ag/AgCl) without the use of mediating redox species and enzyme immobilization by physical adsorption. The biosensor detected as low as 0.5 nM (0.145 ppb) of the model organophosphate nerve agent paraoxon with good precision, electrode to electrode reproducibility and stability. Analysis of real water sample using the sensor demonstrated the feasibility of the application of the sensor for on site monitoring of OP compounds.
引用
收藏
页码:54 / 58
页数:5
相关论文
共 37 条
[31]  
*US DEP AGR, 1992, AGR STAT, P395
[32]   Carbon nanotube screen-printed electrochemical sensors [J].
Wang, J ;
Musameh, M .
ANALYST, 2004, 129 (01) :1-2
[33]   Solubilization of carbon nanotubes by Nafion toward the preparation of amperometric biosensors [J].
Wang, J ;
Musameh, M ;
Lin, YH .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 2003, 125 (09) :2408-2409
[34]  
White PC, 2002, ELECTROANAL, V14, P89, DOI 10.1002/1521-4109(200201)14:2<89::AID-ELAN89>3.0.CO
[35]  
2-Y
[36]   An amperometric biosensor based on the coimmobilization of horseradish peroxidase and methylene blue on a carbon nanotubes modified electrode [J].
Xu, JZ ;
Zhu, JJ ;
Wu, Q ;
Hu, Z ;
Chen, HY .
ELECTROANALYSIS, 2003, 15 (03) :219-224
[37]  
1989, FOOD AGR ORG UN PROD, V43, P320