Transient two-dimensional IR spectrometer for probing nanosecond temperature-jump kinetics
被引:59
作者:
Chung, Hoi Sung
[1
]
Khalil, Munira
Smith, Adam W.
Tokmakoff, Andrei
机构:
[1] MIT, Dept Chem, Cambridge, MA 02139 USA
[2] Univ Washington, Dept Chem, Seattle, WA 98195 USA
基金:
美国国家科学基金会;
关键词:
D O I:
10.1063/1.2743168
中图分类号:
TH7 [仪器、仪表];
学科分类号:
0804 ;
080401 ;
081102 ;
摘要:
We have developed a Fourier transform two-dimensional infrared (2D IR) spectrometer to probe chemical reactions and biophysical processes triggered by a nanosecond temperature jump (T jump). The technical challenges for such a spectrometer involve (1) synchronization of a nanosecond T-jump laser and femtosecond laser system, (2) overcoming the decreased signal-to-noise ratio from low repetition rate data acquisition, and (3) performing an interferometric measurement through a sample with a density and index of refraction that varies with time delay after the T jump. The first challenge was overcome by synchronizing the two lasers to a clock derived from the Ti:sapphire oscillator, leading to timing accuracy of 2 ns for delays up to 50 ms. The data collection time is reduced by using undersampling with the improved signal-to-noise ratio obtained from a balanced detection scheme with a dual stripe array detector. Transient dispersed vibrational echo and 2D IR spectroscopy are applied to N-methylacetamide and ubiquitin, as examples, and the spectral responses by a temperature elevation and by structural changes of the protein are compared. The synchronization of 2D IR spectroscopy with a nanosecond temperature jump without losing its sensitivity at a low repetition rate opens a new applicability of the nonlinear spectroscopy to probe a variety of molecular structure changes induced by a nanosecond perturbation. (c) 2007 American Institute of Physics.
引用
收藏
页数:10
相关论文