摘要
Objectives We performed this study to investigate the vascular response in early period after zortarolimus-eluting stent (ZES) (Endeavor Sprint, Medtronic CardioVascular, Minneapolis, Minnesota) implantation. Background The ZES has different characteristics, with biocompatible polymer and rapid drug-elution, compared with the first-generation drug-eluting stents (DES). Methods The ENDEAVOR OCT (Evaluation in 3 Months Duration of Neointimal Coverage after Zotarolimus-Eluting Stent Implantation by Optical Coherence Tomography) trial is a prospective, single-center study evaluating vascular healing patterns with optical coherence tomography (OCT) at 3 months after stent implantation. A total of 31 ZES in 30 patients underwent serial OCT at immediate post-intervention and 3 months. Neointimal growth and malapposition were analyzed at each stent strut of cross-sectional OCT images with 0.5-mm intervals. Results The incidence of malapposition at post-intervention and 3 months was 6.0% and 0.2%, respectively. However, late acquired malapposition was not detected at 3 months. Of 31 stents, 27 stents (87.1%) were covered completely with neointima, but the remaining 4 stents had 2 (0.8%), 4 (0.9%), 4 (1.2%), and 6 (1.4%) uncovered struts. Overall mean percentage of covered stent struts was 99.9 +/- 0.4%. This finding was consistent among groups with acute coronary syndrome and stable angina pectoris (99.9 +/- 0.3% vs. 99.9 +/- 0.4%, p = 0.92). Intracoronary thrombus was documented in 1 stent (3.2%) among 31 stents. Conclusions Most of the stent struts were covered with neointima, and late acquired malapposition was not found at 3 months after ZES implantation. Therefore, the current study demonstrated that ZES might have a favorable in vivo vascular response at 3 months after stent implantation. (Evaluation of Zotarolimus Eluting Stent at 3 Months Using Optical Coherence Tomography (ENDEAVOR OCT]; NCT00815139) (J Am Coll Cardiol Intv 2009;2:1240-7) (C) 2009 by the American College of Cardiology Foundation
引用
收藏
页码:1240 / 1247
页数:8
相关论文
共 25 条
[1]   Late incomplete stent apposition after sirolimus-eluting stent implantation - A serial intravascular ultrasound analysis [J].
Ako, J ;
Morino, Y ;
Honda, Y ;
Hassan, A ;
Sonoda, S ;
Yock, PG ;
Leon, MB ;
Moses, JW ;
Bonneau, HN ;
Fitzgerald, PJ .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2005, 46 (06) :1002-1005
[2]   Angioscopic comparison of neointimal coverage between zotarolimus- and sirolimus-eluting stents [J].
Awata, Masaki ;
Nanto, Shinsuke ;
Uematsu, Masaaki ;
Morozumi, Takakazu ;
Watanabe, Tetsuya ;
Onishi, Toshinari ;
Iida, Osamu ;
Sera, Fusako ;
Kotani, Jun-ichi ;
Hori, Masatsugu ;
Nagata, Seiki .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2008, 52 (09) :789-790
[3]   A hierarchical Bayesian meta-analysis of randomised clinical trials of drug-eluting stents [J].
Babapulle, MN ;
Joseph, L ;
Bélisle, P ;
Brophy, JM ;
Eisenberg, MJ .
LANCET, 2004, 364 (9434) :583-591
[4]   Predictors of subacute stent thrombosis - Results of a systematic intravascular ultrasound study [J].
Cheneau, E ;
Leborgne, L ;
Mintz, GS ;
Kotani, J ;
Pichard, AD ;
Satler, LF ;
Canos, D ;
Castagna, M ;
Weissman, NJ ;
Waksman, R .
CIRCULATION, 2003, 108 (01) :43-47
[5]   Pathological mechanisms of fatal late coronary stent thrombosis in humans [J].
Farb, A ;
Burke, AP ;
Kolodgie, FD ;
Virmani, R .
CIRCULATION, 2003, 108 (14) :1701-1706
[6]   Pathological correlates of late drug-eluting stent thrombosis - Strut coverage as a marker of endothelialization [J].
Finn, Aloke V. ;
Joner, Michael ;
Nakazawa, Gaku ;
Kolodgie, Frank ;
Newell, John ;
John, Mike C. ;
Gold, Herman K. ;
Virmani, Renu .
CIRCULATION, 2007, 115 (18) :2435-2441
[7]  
GUAGLIUMI G, 2008, LAT BREAK TRIALS TRA
[8]   Late stent malapposition after drug-eluting stent implantation - An intravascular ultrasound analysis with long-term follow-up [J].
Hong, MK ;
Mintz, GS ;
Lee, CW ;
Park, DW ;
Park, KM ;
Lee, BK ;
Kim, YH ;
Song, JM ;
Han, KH ;
Kang, DH ;
Cheong, SS ;
Song, JK ;
Kim, JJ ;
Park, SW ;
Park, SJ .
CIRCULATION, 2006, 113 (03) :414-419
[9]   Visualization of coronary atherosclerotic plaques in patients using optical coherence tomography: Comparison with intravascular ultrasound [J].
Jang, IK ;
Bouma, BE ;
Kang, DH ;
Park, SJ ;
Park, SW ;
Seung, KB ;
Choi, KB ;
Shishkov, M ;
Schlendorf, K ;
Pomerantsev, E ;
Houser, SL ;
Aretz, HT ;
Tearney, GJ .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2002, 39 (04) :604-609
[10]   Stent thrombosis after successful sirolimus-eluting stent implantation [J].
Jeremias, A ;
Sylvia, B ;
Bridges, J ;
Kirtane, AJ ;
Bigelow, B ;
Pinto, DS ;
Ho, KKL ;
Cohen, DJ ;
Garcia, LA ;
Cutlip, DE ;
Carrozza, JP .
CIRCULATION, 2004, 109 (16) :1930-1932