Paper-based microfluidic devices for analysis of clinically relevant analytes present in urine and saliva

被引:250
作者
Klasner, Scott A. [1 ]
Price, Alexander K. [1 ]
Hoeman, Kurt W. [2 ]
Wilson, Rashaun S. [1 ]
Bell, Kayla J. [3 ]
Culbertson, Christopher T. [1 ]
机构
[1] Kansas State Univ, Dept Chem, Manhattan, KS 66506 USA
[2] Halliburton, Duncan Technol Ctr, Duncan, OK 73536 USA
[3] Bethany Coll, Dept Chem, Lindsborg, KS 67456 USA
基金
美国国家科学基金会;
关键词
Paper-based devices; Paper microfluidics; Clinical assays; Colorimetric detection; Photolithography; Photoactive polymers;
D O I
10.1007/s00216-010-3718-4
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
We report the use of paper-based microfluidic devices fabricated from a novel polymer blend for the monitoring of urinary ketones, glucose, and salivary nitrite. Paper-based devices were fabricated via photolithography in less than 3 min and were immediately ready for use for these diagnostically relevant assays. Patterned channels on filter paper as small as 90 mu m wide with barriers as narrow as 250 mu m could be reliably patterned to permit and block fluid wicking, respectively. Colorimetric assays for ketones and nitrite were adapted from the dipstick format to this paper microfluidic chip for the quantification of acetoacetate in artificial urine, as well as nitrite in artificial saliva. Glucose assays were based on those previously demonstrated (Martinez et al., Angew Chem Int Ed 8: 1318-1320, 1; Martinez et al., Anal Chem 10: 3699-3707, 2; Martinez et al., Proc Nat Acad Sci USA 50: 19606-19611, 3; Lu et al., Electrophoresis 9: 1497-1500, 4; Abe et al., Anal Chem 18: 6928-6934, 5). Reagents were spotted on the detection pad of the paper device and allowed to dry prior to spotting of samples. The ketone test was a two-step reaction requiring a derivitization step between the sample spotting pad and the detection pad, thus for the first time, confirming the ability of these paper devices to perform online multi-step chemical reactions. Following the spotting of the reagents and sample solution onto the paper device and subsequent drying, color images of the paper chips were recorded using a flatbed scanner, and images were converted to CMYK format in Adobe Photoshop CS4 where the intensity of the color change was quantified using the same software. The limit of detection (LOD) for acetoacetate in artificial urine was 0.5 mM, while the LOD for salivary nitrite was 5 mu M, placing both of these analytes within the clinically relevant range for these assays. Calibration curves for urinary ketone (5 to 16 mM) and salivary nitrite (5 to 2,000 mu M) were generated. The time of device fabrication to the time of test results was about 25 min.
引用
收藏
页码:1821 / 1829
页数:9
相关论文
共 20 条
[1]   Inkjet-printed microfluidic multianalyte chemical sensing paper [J].
Abe, Koji ;
Suzuki, Koji ;
Citterio, Daniel .
ANALYTICAL CHEMISTRY, 2008, 80 (18) :6928-6934
[2]   Use of colorimetric test strips for monitoring the effect of hemodialysis on salivary nitrite and uric acid in patients with end-stage renal disease: A proof of principle [J].
Blicharz, Timothy M. ;
Rissin, David M. ;
Bowden, Michaela ;
Hayman, Ryan B. ;
DiCesare, Christopher ;
Bhatia, Jasvinder S. ;
Grand-Pierre, Nerline ;
Siqueira, Walter L. ;
Helmerhorst, Eva J. ;
Loscalzo, Joseph ;
Oppenheim, Frank G. ;
Walt, David R. .
CLINICAL CHEMISTRY, 2008, 54 (09) :1473-1480
[3]   A simple artificial urine for the growth of urinary pathogens [J].
Brooks, T ;
Keevil, CW .
LETTERS IN APPLIED MICROBIOLOGY, 1997, 24 (03) :203-206
[4]   Paper Microzone Plates [J].
Carrilho, Emanuel ;
Phillips, Scott T. ;
Vella, Sarah J. ;
Martinez, Andres W. ;
Whitesides, George M. .
ANALYTICAL CHEMISTRY, 2009, 81 (15) :5990-5998
[5]   Understanding Wax Printing: A Simple Micropatterning Process for Paper-Based Microfluidics [J].
Carrilho, Emanuel ;
Martinez, Andres W. ;
Whitesides, George M. .
ANALYTICAL CHEMISTRY, 2009, 81 (16) :7091-7095
[6]   Quantifying Colorimetric Assays in Paper-Based Microfluidic Devices by Measuring the Transmission of Light through Paper [J].
Ellerbee, Audrey K. ;
Phillips, Scott T. ;
Siegel, Adam C. ;
Mirica, Katherine A. ;
Martinez, Andres W. ;
Striehl, Pierre ;
Jain, Nina ;
Prentiss, Mara ;
Whitesides, George M. .
ANALYTICAL CHEMISTRY, 2009, 81 (20) :8447-8452
[7]  
Fischbach FT., 2009, A manual of laboratory and diagnostic tests, V8th
[8]   Formation, collection and significance of gingival crevice fluid [J].
Griffiths, GS .
PERIODONTOLOGY 2000, 2003, 31 :32-42
[9]   KINETIC-STUDY OF THE REACTION OF ACETOACETATE WITH GLYCINE AND SODIUM-NITROPRUSSIDE [J].
LAIOS, ID ;
PARDUE, HL .
ANALYTICAL CHEMISTRY, 1993, 65 (14) :1903-1909
[10]   Paper-Based Microfluidic Devices by Plasma Treatment [J].
Li, Xu ;
Tian, Junfei ;
Nguyen, Thanh ;
Shen, Wei .
ANALYTICAL CHEMISTRY, 2008, 80 (23) :9131-9134