Measuring the diameter of coronary arteries on MR angiograms using spatial profile curves

被引:17
作者
Shimamoto, R
Suzuki, J
Nishikawa, J
Fujimori, Y
Nakamura, F
Shin, WS
Tomaru, T
Toyo-oka, T
机构
[1] Univ Tokyo, Fac Med, Dept Internal Med 2, Bunkyo Ku, Tokyo 1130033, Japan
[2] Univ Tokyo, Fac Med, Dept Radiol, Tokyo 1130033, Japan
关键词
D O I
10.2214/ajr.170.4.9530028
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The spatial profile curve of the nuclear MR intensity across the short axis of a coronary artery in an MR angiogram results in a gradual up-and-down slope lacking sharp definition, which indicates that display parameters may influence edge recognition. Therefore, our study was designed to determine the appropriate window setting and to devise a method of accurately measuring the diameter or width of the artery independent of window parameters. CONCLUSION, The diameter of a coronary artery measured on MR coronary arteriography significantly varied with experimentally selected display parameters, When compared with the diameter on contrast-enhanced coronary arteriograms, the window center on MR angiograms at the midpoint between the peak intensity of the intravascular lumen and the background intensity and the window width of a quarter or a half of the intensity difference between the two were proven to be appropriate. The angiographic diameter corresponded to the diameter obtained at 65% +/- 9% of the peak intensity on the spatial profile curve across the short-axis MR coronary angiogram. Accordingly, 65% of the peak intensity indicates the diameter of the coronary artery, Thus, the intensity profile curve independent of the window setting provided a new method for measurement of the diameter of the coronary artery.
引用
收藏
页码:889 / 893
页数:5
相关论文
共 9 条
[1]   CINEANGIOGRAPHY OF THE HEART IN A SINGLE BREATH HOLD WITH A SEGMENTED TURBOFLASH SEQUENCE [J].
ATKINSON, DJ ;
EDELMAN, RR .
RADIOLOGY, 1991, 178 (02) :357-360
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   2-DIMENSIONAL CORONARY MR-ANGIOGRAPHY - ANALYSIS OF INITIAL CLINICAL-RESULTS [J].
DUERINCKX, AJ ;
URMAN, MK .
RADIOLOGY, 1994, 193 (03) :731-738
[4]   CORONARY-ARTERIES - BREATH-HOLD MR ANGIOGRAPHY [J].
EDELMAN, RR ;
MANNING, WJ ;
BURSTEIN, D ;
PAULIN, S .
RADIOLOGY, 1991, 181 (03) :641-643
[5]   AN IMPROVED QUADRATURE OR PHASED-ARRAY COIL FOR MR CARDIAC IMAGING [J].
FAYAD, ZA ;
CONNICK, TJ ;
AXEL, L .
MAGNETIC RESONANCE IN MEDICINE, 1995, 34 (02) :186-193
[6]   A PRELIMINARY-REPORT COMPARING MAGNETIC-RESONANCE CORONARY ANGIOGRAPHY WITH CONVENTIONAL ANGIOGRAPHY [J].
MANNING, WJ ;
LI, W ;
EDELMAN, RR .
NEW ENGLAND JOURNAL OF MEDICINE, 1993, 328 (12) :828-832
[7]   FAT-SUPPRESSED BREATH-HOLD MAGNETIC-RESONANCE CORONARY ANGIOGRAPHY [J].
MANNING, WJ ;
LI, W ;
BOYLE, NG ;
EDELMAN, RR .
CIRCULATION, 1993, 87 (01) :94-104
[8]  
PENNELL DJ, 1993, BRIT HEART J, V70, P315
[9]   MAGNETIC-RESONANCE ANGIOGRAPHY OF ANOMALOUS CORONARY-ARTERIES - A NEW GOLD STANDARD FOR DELINEATING THE PROXIMAL COURSE [J].
POST, JC ;
VANROSSUM, AC ;
BRONZWAER, JGF ;
DECOCK, CC ;
HOFMAN, MBM ;
VALK, J ;
VISSER, CA .