Development of a sample equilibration system for the TEOM continuous PM monitor

被引:51
作者
Meyer, MB [1 ]
Patashnick, H [1 ]
Ambs, JL [1 ]
Rupprecht, E [1 ]
机构
[1] Rupprecht & Patashnick Co Inc R&P, Albany, NY 12203 USA
来源
JOURNAL OF THE AIR & WASTE MANAGEMENT ASSOCIATION | 2000年 / 50卷 / 08期
关键词
D O I
10.1080/10473289.2000.10464180
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
In recent years, scientific discussion has included the influence of thermodynamic conditions (e.g., temperature, relative humidity, and filter face velocity) on PM retention efficiency of filter-based samplers and monitors. Method-associated thermodynamic conditions can, in some instances, dramatically influence the presence of particle-bound water and other light-molecular-weight chemical components such as particulate nitrates and certain organic compounds. The measurement of fine particle mass presents a new challenge for all PM measurement methods, since a relatively greater fraction of the mass is semi-volatile. The tapered element oscillating microbalance (TEOM) continuous PM monitor is a U.S. Environmental Protection Agency (EPA) PM10 equivalent method (EQPM-1090079). Several hundred of these monitors are deployed throughout the United States. The TEOM monitor has the unique characteristic of providing direct PM mass measurement without the calibration uncertainty inherent in mass surrogate methods. In addition, it provides high-precision, near-real-time continuous data automatically. Much attention has been given to semi-volatile species retention of the TEOM method. While using this monitor, it is desirable to maintain as low an operating temperature as practical and to remove unwanted particle-bound water. A new sample equilibration system (SES) has been developed to allow conditioning of the PM sample stream to a lower humidity and temperature level. The SES incorporates a special low-particle-loss Nafion dryer. This paper discusses the configuration and theory of the SES. Performance results include high time-resolved PM2.5 data comparison between a 30 degrees C sample stream TEOM monitor with SES and a standard 50 degrees C TEOM monitor. In addition, 24-hr integrated data are compared with data collected using an EPA PM2.5 Federal Reference Method (FRM)-type sampler. The SES is a significant development because it can be applied easily to existing TEOM monitors.
引用
收藏
页码:1345 / 1349
页数:5
相关论文
共 10 条
  • [1] Evaluation of the TEOM(R) method for measurement of ambient particulate mass in urban areas
    Allen, G
    Sioutas, C
    Koutrakis, P
    Reiss, R
    Lurmann, FW
    Roberts, PT
    [J]. JOURNAL OF THE AIR & WASTE MANAGEMENT ASSOCIATION, 1997, 47 (06) : 682 - 689
  • [2] [Anonymous], FED REG
  • [3] TEOM vs. manual gravimetric methods for determination of PM2.5 aerosol mass concentrations
    Ayers, GP
    Keywood, MD
    Gras, JL
    [J]. ATMOSPHERIC ENVIRONMENT, 1999, 33 (22) : 3717 - 3721
  • [4] Dick W., 1995, PTL PUBLICATION
  • [5] Integrated and real-time diffusion denuder sampler for PM2.5
    Eatough, DJ
    Obeidi, F
    Pang, YB
    Ding, YM
    Eatough, NL
    Wilson, WE
    [J]. ATMOSPHERIC ENVIRONMENT, 1999, 33 (17) : 2835 - 2844
  • [6] Effects of equilibration temperature on PM10 concentrations from the TEOM method in the Lower Fraser Valley
    Mignacca, D
    Stubbs, K
    [J]. JOURNAL OF THE AIR & WASTE MANAGEMENT ASSOCIATION, 1999, 49 (10): : 1250 - 1254
  • [7] CONTINUOUS PM-10 MEASUREMENTS USING THE TAPERED ELEMENT OSCILLATING MICROBALANCE
    PATASHNICK, H
    RUPPRECHT, EG
    [J]. JOURNAL OF THE AIR & WASTE MANAGEMENT ASSOCIATION, 1991, 41 (08): : 1079 - 1083
  • [8] Patashnick H, 1998, J AIR WASTE MANAGE, V48, P195
  • [9] PATASHNICK H, IN PRESS AEROSOL SCI
  • [10] 1999, GEMEINSAMES MINISTER, V22, P448