Interference due to haemolysis in routine photometric analysis - a survey

被引:11
作者
Brady, J [1 ]
O'Leary, N
机构
[1] Our Ladys Hosp Sick Children, Dept Clin Biochem, Dublin 12, Ireland
[2] Natl Univ Ireland Univ Coll Dublin, Dept Biochem, Dublin 4, Ireland
[3] Cork Univ Hosp, Dept Biochem, Cork, Ireland
关键词
external quality assessment; haemoglobin; spectrophotometry;
D O I
10.1177/000456329803500118
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
To assess the affect of haemolysis on routine photometric methods used by participants in the Irish External Quality Assessment Scheme for Clinical Biochemistry (IEQAS) a sample was distributed with a supplied diluent containing haemoglobin. A sample of the same batch had previously been circulated as a normal distribution. Many analytes/methods were affected by the haemolysis. The range of values reported for each analyte was as follows. Albumin 34-48 g/L (baseline 37 g/L). Calcium 2.16-2.68 mmol/L (baseline 2.34 mmol/L). Urate 90-470 mu mol/L (baseline 195 mu mol/L). Creatinine 55-161 mu mol/L (baseline 140 mu mol/L). Glucose 4.1- 5-6 mmol/L (baseline 4.8 mmol/L), Bilirubin 14-60 mu mol/L (baseline 36 mu mol/L). Urea 3.4-7.3 mmol/L (baseline 3.8 mmol/L). Iron 17.9-55.0 mu mol/L (baseline 26.8 mu mol/L). Inorganic phosphorus 0.22-2.41 mmol/L (baseline 1.59 mmol/L). Cholesterol 3.0-5.9 mmol/L (baseline 4.1 mmol/L). In some situations the use of an appropriate blank may solve the problem, but it must be remembered that interference caused by haemoglobin may be more complex than merely spectral.
引用
收藏
页码:128 / 134
页数:7
相关论文
共 11 条
[11]  
SCHLEBUSCH H, 1993, CLIN CHEM, V39, P1993