Movable reduction bed hydride generation system as an interface for capillary zone electrophoresis and inductively coupled plasma atomic emission spectrometry for arsenic speciation analysis

被引:47
作者
Tian, XD [1 ]
Zhuang, ZX [1 ]
Chen, B [1 ]
Wang, XR [1 ]
机构
[1] Xiamen Univ, Dept Chem, Lab Analyt Sci Mat & Life Proc, Xiamen 361005, Peoples R China
关键词
interface; moveable reduction bed hydride generator; capillary zone electrophoresis; inductively coupled plasma; arsenic; speciation analysis;
D O I
10.1039/a707452b
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Hydride generation can significantly improve the detection limits for a number of hydride-forming elements with atomic spectrometric determination, In order to harmonize the microliter level eluates from capillary electrophoresis and the milliliter sample volume commonly needed in ICP-AES, a specially designed technique, 'movable reduction bed hydride generation? has been developed, which is simple to operate and carries out the hydride generation for micro-sample volume without the use of a gas-liquid separator or the addition of an acid dehydrator. The technique can be used to convert arsenic compounds from capillary zone electrophoresis (CZE) separation to corresponding volatile hydrides followed by ICP-AES analysis. The separation characteristics of arsenic species by CZE are presented, including the optimization of experimental parameters such as pH, concentration of buffer solution, as well as sample injection volume. The RSDs of peak area for four arsenic species range from 1.9 to 11.7%, The detection limits for As-III, disodium methylarsonate and As-V are 0.32 mu g ml(-1), and 0.35 mu g ml(-1) for dimethylarsinic acid.
引用
收藏
页码:899 / 903
页数:5
相关论文
共 28 条
[1]   ANALYSIS OF ORGANIC AND NONORGANIC ARSENIOUS OR SELENIOUS COMPOUNDS BY CAPILLARY ELECTROPHORESIS [J].
ALBERT, M ;
DEMESMAY, C ;
ROCCA, JL .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1995, 351 (4-5) :426-432
[2]   HIGH-VOLTAGE CAPILLARY ZONE ELECTROPHORESIS - OPERATING PARAMETERS EFFECTS ON ELECTROENDOSMOTIC FLOWS AND ELECTROPHORETIC MOBILITIES [J].
ALTRIA, KD ;
SIMPSON, CF .
CHROMATOGRAPHIA, 1987, 24 :527-532
[3]  
ALTRIA KD, 1988, ANAL P, V25, P85
[4]  
ANDRAE MO, 1986, ORGANOMETALLIC COMPO
[5]   ARSENIC SPECIATION BY COUPLING HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY WITH INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY [J].
DEMESMAY, C ;
OLLE, M ;
PORTHAULT, M .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1994, 348 (03) :205-210
[6]   CAPILLARY ELECTROPHORESIS [J].
EWING, AG ;
WALLINGFORD, RA ;
OLEFIROWICZ, TM .
ANALYTICAL CHEMISTRY, 1989, 61 (04) :A292-+
[7]   SPECIATION OF ARSENIC IN URINE USING HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY WITH INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRIC DETECTION [J].
HEITKEMPER, D ;
CREED, J ;
CARUSO, J .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1989, 4 (03) :279-284
[8]   CAPILLARY ZONE ELECTROPHORESIS [J].
JORGENSON, JW ;
LUKACS, KD .
SCIENCE, 1983, 222 (4621) :266-272
[9]   ZONE ELECTROPHORESIS IN OPEN-TUBULAR GLASS-CAPILLARIES [J].
JORGENSON, JW ;
LUKACS, KD .
ANALYTICAL CHEMISTRY, 1981, 53 (08) :1298-1302
[10]   PH HYSTERESIS EFFECT WITH SILICA CAPILLARIES IN CAPILLARY ZONE ELECTROPHORESIS [J].
LAMBERT, WJ ;
MIDDLETON, DL .
ANALYTICAL CHEMISTRY, 1990, 62 (15) :1585-1587