Reproducibility of optic disk topographic measurements with the Topcon ImageNet and the Heidelberg retina tomograph

被引:16
作者
Azuara-Blanco, A
Harris, A
Cantor, LB
机构
[1] Indiana Univ, Sch Med, Dept Ophthalmol, Indianapolis, IN 46202 USA
[2] Indiana Univ, Sch Med, Dept Physiol & Biophys, Indianapolis, IN 46202 USA
[3] Thomas Jefferson Univ, Jefferson Med Coll, Wills Eye Hosp, Glaucoma Serv, Philadelphia, PA 19107 USA
关键词
optic nerve head; reliability; variability; diagnosis; image analysis; laser; confocal laser;
D O I
10.1159/000027286
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Objective: To compare the reproducibility of optic disk measurements provided by an image analyzer and a scanning laser tomograph. Methods: Ten images of the same eye of 10 normal volunteers were taken with the Heidelberg Retina Tomograph and with the Topcon ImageNet. Intraclass correlation coefficient (ICC) and coefficient of variation (CV) were used to evaluate the reproducibility of the measurements. Results: Eleven parameters were analyzed with the Topcon ImageNet. Six parameters (55%) had ICC greater than 90%. Four parameters (36%) had CV less than 10%. Twelve parameters were evaluated with the Heidelberg Retina Tomograph. Nine parameters (75%) had ICC over 90%. Nine parameters (75%) had CV less than 10%. Conclusion: Both systems provided reproducible data. The optic disk parameters provided by the Heidelberg Retina Tomograph had a better reproducibility than those obtained from the Topcon ImageNet.
引用
收藏
页码:95 / 98
页数:4
相关论文
共 22 条
  • [21] EFFECT OF REPETITIVE IMAGING ON TOPOGRAPHIC MEASUREMENTS OF THE OPTIC-NERVE HEAD
    WEINREB, RN
    LUSKY, M
    BARTSCH, DU
    MORSMAN, D
    [J]. ARCHIVES OF OPHTHALMOLOGY, 1993, 111 (05) : 636 - 638
  • [22] WINER BJ, 1971, STATISTICAL PRINCIPL, P244