Complexity-measure-based sequential hypothesis testing for real-time detection of lethal cardiac arrhythmias

被引:10
作者
Chen, Szi-Wen [1 ]
机构
[1] Chang Gung Univ, Dept Elect Engn, Tao Yuan 333, Taiwan
关键词
D O I
10.1155/2007/20957
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A novel approach that employs a complexity-based sequential hypothesis testing (SHT) technique for real-time detection of ventricular fibrillation (VF) and ventricular tachycardia (VT) is presented. A dataset consisting of a number of VF and VT electrocardiogram ( ECG) recordings drawn from the MIT-BIH database was adopted for such an analysis. It was split into two smaller datasets for algorithm training and testing, respectively. Each ECG recording was measured in a 10-second interval. For each recording, a number of overlapping windowed ECG data segments were obtained by shifting a 5-second window by a step of 1 second. During the windowing process, the complexity measure (CM) value was calculated for each windowed segment and the task of pattern recognition was then sequentially performed by the SHT procedure. A preliminary test conducted using the database produced optimal overall predictive accuracy of 96.67%. The algorithm was also implemented on a commercial embedded DSP controller, permitting a hardware realization of real-time ventricular arrhythmia detection.