Effect of oxidation on delamination of ultrahigh-molecular-weight polyethylene tibial components

被引:82
作者
Bell, CJ [1 ]
Walker, PS [1 ]
Abeysundera, MR [1 ]
Simmons, JMH [1 ]
King, PM [1 ]
Blunn, GW [1 ]
机构
[1] UCL, Royal Natl Orthopaed Hosp Trust, Ctr Biomed Engn, Stanmore HA7 4LP, Middx, England
关键词
ultrahigh-molecular-weight polyethylene; total knee arthroplasty; delamination; oxidation;
D O I
10.1016/S0883-5403(98)90173-5
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Whether oxidation of ultrahigh-molecular-weight polyethylene (UHMWPE) causes delamination of the plastic in total knee arthroplasties (TKAs) was investigated. Examination of retrieved TKAs has shown that oxidation of UHMWPE can be caused by postirradiation damage leading to a subsurface band or, to a lesser extent, by mechanical forces during use leading to surface oxidation. Delamination cracks propagated through the subsurface oxidized band. In wear tests, delamination occurred in artificially aged UHMWPE where only subsurface oxidized bands had formed. Increased surface wear predominated where oxidation was associated with the surface of the plastic. Similarly, in tensile and fatigue tests of oxidized UHMWPE, there was a significant reduction in the ultimate tensile strength and in the fatigue resistance of specimens that had developed a subsurface band. Oxidation increased fatigue crack growth rate. It was observed that UHMWPE from different manufacturers varied in its resistance to oxidation. This study demonstrates that the effect of oxidation, which results in the development of a subsurface white band, combined with high subsurface shear forces observed in TKAs, is to enhance delamination wear.
引用
收藏
页码:280 / 290
页数:11
相关论文
共 24 条
[1]   THE EFFECT OF CONFORMITY, THICKNESS, AND MATERIAL ON STRESSES IN ULTRAHIGH MOLECULAR-WEIGHT COMPONENTS FOR TOTAL JOINT REPLACEMENT [J].
BARTEL, DL ;
BICKNELL, VL ;
WRIGHT, TM .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1986, 68A (07) :1041-1051
[2]   POLYETHYLENE WEAR IN UNICONDYLAR KNEE PROSTHESES - 106 RETRIEVED MARMOR, PCA, AND ST-GEORG TIBIAL COMPONENTS COMPARED [J].
BLUNN, GW ;
JOSHI, AB ;
LILLEY, PA ;
ENGELBRECHT, E ;
RYD, L ;
LIDGREN, L ;
HARDINGE, K ;
NIEDER, E ;
WALKER, PS .
ACTA ORTHOPAEDICA SCANDINAVICA, 1992, 63 (03) :247-255
[3]   Wear in retrieved condylar knee arthroplasties - A comparison of wear in different designs of 280 retrieved condylar knee prostheses [J].
Blunn, GW ;
Joshi, AB ;
Minns, RJ ;
Lidgren, L ;
Lilley, P ;
Ryd, L ;
Engelbrecht, E ;
Walker, PS .
JOURNAL OF ARTHROPLASTY, 1997, 12 (03) :281-290
[4]  
BLUNN GW, 1991, CLIN ORTHOP RELAT R, P253
[5]  
Chapiro A., 1974, Technical developments and prospects of sterilization by ionizing radiation, P367
[6]  
CONNELLY G M, 1984, Journal of Orthopaedic Research, V2, P119, DOI 10.1002/jor.1100020202
[7]   PRELIMINARY-STUDY OF THE EFFECT OF AGING FOLLOWING IRRADIATION ON THE WEAR OF ULTRAHIGH-MOLECULAR-WEIGHT POLYETHYLENE [J].
FISHER, J ;
CHAN, KL ;
HAILEY, JL ;
SHAW, D ;
STONE, M .
JOURNAL OF ARTHROPLASTY, 1995, 10 (05) :689-692
[8]  
FISHER J, 1995, T ORTHOP RES SOC, V20, P120
[9]  
GUEUGNAUT D, 1988, J APPL POLYM SCI, V35, P1693
[10]  
HAMILTON JV, 1996, T ORTHOP RES SOC, V21, P20