Validation of the single-platform ISHAGE method for CD34+ hematopoietic stem and progenitor cell enumeration in an international multicenter study

被引:107
作者
Gratama, JW
Kraan, J
Keeney, M
Sutherland, DR
Granger, V
Barnett, D
机构
[1] Dr Daniel Den Hoed Canc Ctr, Lab Clin & Tumor Immunol, Erasmus Med Ctr, Dept Internal Oncol, NL-3075 EA Rotterdam, Netherlands
[2] London Hlth Sci Ctr, Dept Hematol, London, ON, Canada
[3] Univ Hlth Network, Princess Margaret Hosp, Toronto, ON, Canada
[4] Royal Hallamshire Hosp, Dept Haematol, UK NEQAS Leukocyte Immunophenotyping, Sheffield S10 2JF, S Yorkshire, England
关键词
flow cytometry; CD34; hematopoietic stem and progenitor cells; multicenter study; variation; standardization;
D O I
10.1080/14653240310000083
中图分类号
Q813 [细胞工程];
学科分类号
摘要
Background Flow cytometric enumeration of CD34(+) hematopoietic stem and progenitor cells (HPC) is the reference point for undertaking apheresis and evaluation of adequacy for PBSC engraftment. An external quality assurance (EQA) scheme for CD34(+) HPC enumeration has been operational in Belgium, Netherlands and Luxemburg (Benelux) since 1995. Within this group, a multicenter survey was held to validate the state-of-the-art methodology, i.e., multiparametric definition of HPC based on light scatter, expression of CD34 and CD45, and counting beads (i.e., 'single platform ISHAGE' method). Methods 'Real-time' EQA was used to monitor the application of the single-platform ISHAGE method by 36 participants. Three send-outs of stabilized blood with CD34(+) cell counts 35-60 cells/mL were distributed to 36 participants, who were required to assay the samples on three occasions using the standard assay and their local techniques. These results were compared with those obtained by 111-116 UK NEQAS participants testing the same specimens. Results Using the single platform ISHAGE method, between-laboratory coefficients of variation (CVs) as low as 10% were achieved. Intra-laboratory CVs were < 5% for similar to 50% of the participants. Local single-platform techniques yielded between-laboratory CVs as low as 9% in both Benelux and UK NEQAS cohorts. In contrast, the lowest between-laboratory CVs using dual-platform techniques were 17% (Benelux) and 21% (UK NEQAS), respectively. Conclusion The single-platform ISHAGE method for CD34(+) cell enumeration has been validated by an international group of 36 laboratories. The observed variation between laboratories allows a meaningful comparison of CD34(+) cell enumeration results.
引用
收藏
页码:55 / 65
页数:11
相关论文
共 23 条
[1]   Number of viable CD34+ cells reinfused predicts engraftment in autologous hematopoietic stem cell transplantation [J].
Allan, DS ;
Keeney, M ;
Howson-Jan, K ;
Popma, J ;
Weir, K ;
Bhatia, M ;
Sutherland, DR ;
Chin-Yee, IH .
BONE MARROW TRANSPLANTATION, 2002, 29 (12) :967-972
[2]  
BARNETT D, 2003, BRIT J HAEMATOL, V106, P1059
[3]  
BARNETT D, 2003, BRIT J HAEMATOL, V102, P553
[4]  
BARNETT D, 2003, BRIT J HAEMATOL, V108, P784
[5]  
BARNETT D, 1998, Patent No. 2279653
[6]  
BENDER JG, 2003, HEMATOPOIETIC STEM C, P31
[7]   ANTIGEN CD34+ MARROW-CELLS ENGRAFT LETHALLY IRRADIATED BABOONS [J].
BERENSON, RJ ;
ANDREWS, RG ;
BENSINGER, WI ;
KALAMASZ, D ;
KNITTER, G ;
BUCKNER, CD ;
BERNSTEIN, ID .
JOURNAL OF CLINICAL INVESTIGATION, 1988, 81 (03) :951-955
[8]  
Brando B, 2000, CYTOMETRY, V42, P327, DOI 10.1002/1097-0320(20001215)42:6<327::AID-CYTO1000>3.0.CO
[9]  
2-F
[10]   Enumeration of CD34+ cells in cord blood: A variation on a single-platform flow cytometric method based on the ISHAGE gating strategy [J].
Brocklebank, AM ;
Sparrow, RL .
CYTOMETRY, 2001, 46 (04) :254-261