Modified sector-integration method for predicting the output factors of electron beams including extended source to surface distance

被引:10
作者
Choi, DR
Wolters, J
Mason, D
Baillie, A
机构
[1] Univ Calgary, Tom Baker Canc Ctr, Dept Med Phys, Calgary, AB T2N 4N2, Canada
[2] Univ Calgary, Dept Oncol, Calgary, AB T2N 4N2, Canada
[3] British Columbia Canc Agcy, Dept Med Phys, Canc Ctr So Interior, Kelowna, BC V1Y 5L3, Canada
关键词
D O I
10.1088/0031-9155/45/11/318
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A modified sector-integration method is presented that can predict the output factors of irregular shaped electron fields even in the case of extended source to surface distance (SSD). The model takes as input measured output factors for circular inserts of various radii. These circular fields were measured at SSDs of 100, 105 and 110 cm to determine the effective source distance as a function of radius (ESD(r)). For an arbitrary electron field at any SSD, the shape is divided into small sectors, and the contribution calculated from the radius and ESD(r). The calculated output factors were verified by direct measurements of various types of electron fields mainly based on clinical use. The energies modelled were 8, 10 and 12 MeV for applicator sizes of 10 cm x 10 cm and 14 cm x 14 cm (defined at 95 cm). The calculated values agreed with the measured data within 1% for the various rectangular cutouts including extended source to surface distance. We retrospectively modelled 97 patient inserts of irregular shape, and found agreement within 2% of measured values.
引用
收藏
页码:3367 / 3372
页数:6
相关论文
共 15 条
[1]  
ALGHAZI MSA, 1993, MED PHYS, V20, P886
[2]   ELECTRON DOSIMETRY OF IRREGULAR FIELDS ON THE CLINAC-18 [J].
BIGGS, PJ ;
BOYER, AL ;
DOPPKE, KP .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1979, 5 (03) :433-440
[3]   CALCULATION OF ELECTRON-BEAM DEPTH-DOSE CURVES AND OUTPUT FACTORS FOR ARBITRARY FIELD SHAPES [J].
BRUINVIS, IAD ;
MATHOL, WAF .
RADIOTHERAPY AND ONCOLOGY, 1988, 11 (04) :395-404
[4]   Practical approach to electron beam dosimetry at extended SSD [J].
Cygler, J ;
Li, XA ;
Ding, GX ;
Lawrence, E .
PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (08) :1505-1514
[5]   A METHOD FOR THE ASSESSMENT OF THE OUTPUT OF IRREGULARLY SHAPED ELECTRON FIELDS [J].
JONES, D ;
ANDRE, P ;
WASHINGTON, JT ;
HAFERMANN, MD .
BRITISH JOURNAL OF RADIOLOGY, 1990, 63 (745) :59-64
[6]   A sector-integration method for calculating the output factors of irregularly shaped electron fields [J].
Jursinic, PA ;
Mueller, R .
MEDICAL PHYSICS, 1997, 24 (11) :1765-1769
[7]  
KHAN FM, 1994, PHYSICS RAD THERAPY
[8]   A PARAMETERIZATION OF THE ELECTRON-BEAM OUTPUT FACTORS OF A 25-MEV LINEAR-ACCELERATOR [J].
MCPARLAND, BJ .
MEDICAL PHYSICS, 1987, 14 (04) :665-669
[9]   A METHOD OF CALCULATING THE OUTPUT FACTORS OF ARBITRARILY SHAPED ELECTRON FIELDS [J].
MCPARLAND, BJ .
MEDICAL PHYSICS, 1989, 16 (01) :88-93
[10]   PREDICTION OF ELECTRON-BEAM OUTPUT FACTORS [J].
MILLS, MD ;
HOGSTROM, KR ;
ALMOND, PR .
MEDICAL PHYSICS, 1982, 9 (01) :60-68